R E M O T E V O L T A G E S E N S I N G A T T H E R E A R P A N E - Agilent Technologies E3633A Operating Manual

Hide thumbs Also See for E3633A:
Table of Contents

Advertisement

Chapter 3 Front-Panel Operation
Remote Voltage Sensing at the Front and Rear Terminals
Figure 3-2. Remote Voltage Sensing Connections
Remote Voltage Sensing at the Rear Panel
E x t e r n a l s e n s e t e r m i n a l s a r e a l s o a v a i l a b l e o n t h e b a c k o f t h e p o w e r s u p p l y
t h a t a l l o w t h e r e a r o u t p u t v o l t a g e s t o b e s e n s e d a t t h e l o a d , w h i c h c o m p e n s a t e s
f o r i m p e d a n c e l o s s e s i n t h e l o a d w i r i n g . T h e f r o n t p a n e l b i n d i n g p o s t s a r e
p a r a l l e l e d w i t h t h e r e a r o u t p u t t e r m i n a l s .
T h e r e a r o u t p u t t e r m i n a l s a c c e p t w i r e s s i z e s f r o m A W G 2 2 t o A W G 1 0 .
T o m i n i m i z e t h e p o s s i b i l i t y o f i n s t a b i l i t y o n t h e o u t p u t , k e e p l o a d l e a d s a s s h o r t
a s p o s s i b l e a n d b u n d l e o r t w i s t t h e l e a d s t i g h t l y t o g e t h e r t o m i n i m i z e
i n d u c t a n c e .
N o t e
For rear local voltage sensing connections, the front shorting bars must be removed
first and connect the sense wires as shown in Figure 3-3.
Figure 3-3. Rear local sensing connections
3
39

Advertisement

Table of Contents
loading

This manual is also suitable for:

E3634a

Table of Contents