Chapter 8 Specifications
Performance Specifications
Transient Response Time
μ
L e s s t h a n 5 0
s e c f o r o u t p u t t o r e c o v e r t o w i t h i n 1 5 m V f o l l o w i n g a c h a n g e i n
o u t p u t c u r r e n t f r o m f u l l l o a d t o h a l f l o a d o r v i c e v e r s a

Command Processing Time

A v e r a g e t i m e f o r o u t p u t v o l t a g e t o b e g i n t o c h a n g e a f t e r r e c e i p t o f d i g i t a l d a t a
w h e n t h e p o w e r s u p p l y i s c o n n e c t e d d i r e c t l y t o t h e G P I B o r R S - 2 3 2 i s l e s s t h a n
1 0 0 m s e c
OVP and OCP Accuracy
, ± ( % o f o u t p u t + o f f s e t )
O V P
0 . 5 % + 0 . 5 V
O C P
0 . 5 % + 0 . 5 A
Activation time
: A v e r a g e t i m e f o r o u t p u t t o s t a r t t o d r o p a f t e r O V P o r O C P
c o n d i t i o n o c c u r s .
OVP
< 1 . 5 m s e c w h e n t h e t r i p v o l t a g e i s e q u a l o r g r e a t e r t h a n 3 V
< 1 0 m s e c w h e n t h e t r i p v o l t a g e i s l e s s t h a n 3 V
OCP
< 1 0 m s e c