Suffixes And Multipliers; Response Data Types; Scpi Command Completion - Agilent Technologies 66111A User Manual

Fast transient dc source/mobile communications dc source
Table of Contents

Advertisement

Suffixes and Multipliers

Class
Current
Amplitude
Time

Response Data Types

Character strings returned by query statements may take either of the following forms, depending on the
length of the returned string:
<CRD>
Character Response Data. Permits the return of character strings.
<AARD>
Arbitrary ASCII Response Data. Permits the return of undelimited 7-bit ASCII. This data
type has an implied message terminator.
<SRD>
String Response Data. Returns string parameters enclosed in double quotes.

SCPI Command Completion

SCPI commands sent to the dc source are processed either sequentially or in parallel. Sequential
commands finish execution before a subsequent command begins. Parallel commands allow other
commands to begin executing while the parallel command is still executing. Commands that affect
trigger actions are among the parallel commands.
The *WAI, *OPC, and *OPC? common commands provide different ways of indicating when all
transmitted commands, including any parallel ones, have completed their operations. The syntax and
parameters for these commands are described in chapter 8. Some practical considerations for using these
commands are as follows:
*WAI
This prevents the dc source from processing subsequent commands until all pending
operations are completed.
*OPC?
This places a 1 in the Output Queue when all pending operations have completed.
Because it requires your program to read the returned value before executing the next
program statement, *OPC? can be used to cause the controller to wait for commands to
complete before proceeding with its program.
*OPC
This sets the OPC status bit when all pending operations have completed. Since your
program can read this status bit on an interrupt basis, *OPC allows subsequent
commands to be executed.
NOTE:
The trigger subsystem must be in the Idle state for the status OPC bit to be true. As far as
triggers are concerned, OPC is false whenever the trigger subsystem is in the Initiated
state.
Suffix
Unit
A
ampere
V
volt
S
second
Common Multipliers
1E3
K
1E-3
M
1E-6
U
Introduction to Programming - 6
Unit with Multiplier
MA (milliampere)
MV (millivolt)
MS (millisecond)
kilo
milli
micro
73