The grounding cable plug slips out of the jack, or there is a loose connection
Streaks and "scratches" on the display foil?
There are very many small TXT-files on the SD-card?
An existing audio note it is not shown in the diagram
window of NFAsoft.
© Gigahertz Solutions GmbH
The plug is inserted too far
into the isolating material, so
there is no connection
WRONG
In order to assure a high resolution potential free measurement
of the E-field, a highly sensitive foil of a vapour deposited metal
layer has had to be mounted onto the display. Unfortunately,
scratches and streaks cannot be completely avoided.
Solution 1: You have activated the function "permanent log-
ging", which causes the meter to log not only when switched to
"Log". This option will cause a new file to be started each time
the meter is switched on, and each time you switch from one
Solution 2: There is a file system error on your SD-card. In this
case, save the "good" files and re-format your SD-card ("quick
Note: Windows XP does NOT format the SD memory cards according to the
standard (Vista or Windows 7 are ok). XP-users may use the following link to down-
http://www.sdcard.org/consumers/formatter_3/SDFormatterEN2905.zip
The audio files (.wav) will only be displayed in the diagram if
they are stored in the same folder as the LOG-file and if they
Solution: Hold on to the middle
part of the contact on the plug
with a tong, and screw back the
isolation material
Correct
format" or simple deletion is not enough).
load a standard-conformed formatter:
- 31 -
350-027_NFA1000400-6-2_EN-von-susanne as of October 2011
mode to the other.
are not renamed.