Viewing Basic Mode Results:
Press M to turn your meter on. Press M again to view
the memory.
The 14 day average will
appear in the center of the
display with the number of
results shown at the top of
the display. Any control
results are not included in
this 14 day average.
Press ▲ to view the 7 day
high and low summary.
At the top of the display you
will see the number of HI re-
sults (above 10.0 mmol/L) and
LO results (below 4.0 mmol/L)
during the preceding 7 day
period and the total number
of results during that period.
Press ▼ two times to
review previous results.
Results will be shown starting
with the most recent. Each
result will show the date and
time the test was taken.
You can press ▲ or ▼ to
scroll forward or backward
through the results.
30
Need Help? Bayer Diabetes Care Helpline: 1-800-268-7200
When
appears on the
display, you have viewed all of
the results in the memory.
NOTES:
Averages include the specified time period plus the
current day.
The meter will hold 480 results in the memory. When the
maximum is reached, the oldest test result will be removed
as a new test is completed and saved to the memory.
Press M to turn the meter off, or it will automatically turn off
after 3 minutes.
represents the number of test results.