Maintenance and
storage
10
• For cleaning the monitor only use a soft, lightly
moistened cloth.
Do not use petroleum spirits, thinners or similar
solvents!
• Stains on the cuff can be carefully removed using
a moist cloth and soapy water.
Do not wash the cuff or make it wet!
• Protect your monitor against vibrations and do
not leave it in a place where temperatures are
extremely low (below -20 °C) or extremely high
(above 60 °C) or with high humidity (+ 85%).
• Do not use your monitor at very low temperatu-
res (below 10 °C) or very high temperatures
(above 40 °C).
• Do not fold the cuff or air tube together too
tightly and do not crease them.
• Do not carry out repairs of any kind yourself.
If a defect occurs, consult the OMRON
Customer Service or seek the advice of your
surgical appliance stockist or pharmacist.
• The well confirmed accuracy of this blood
pressure monitor is designed to last a long life.
• In order to increase the lifetime of the battery
take it out when you store the device for a
longer period (2 months).
• It is generally recommended to reconfirm the
function and accuracy every 2 years by your
OMRON distributor (address see attached
literature).

Technical data

Model:
Display:
Measurement Range:
Accuracy/Calibration:
Inflation:
Deflation:
Rapid Pressure Release:
Pressure Detection:
Power Source:
Battery Life:
Operating Temperature/
Humidity:
Storage Temperature:
Console Weight:
Outer Dimensions:
Cuff Range:
Accessories:
Note: These specifications, to improve performance, are subject to change without notice
Manufacturer:
This device fulfils the provisions of the EC
directive 93/42/EEC (Medical Device Directive).
This blood pressure monitor is designed according
to the european standard EN1060, Non-invasive
sphygmomanometers Part 1: General requirements
and Part 3: Supplementary requirements for elec-
tromechanical blood pressure measuring systems.
OMRON M1
LCD digital display
Pressure: 0 to 280 mmHg
Pulse: 40 – 200/min.
Pressure: ± 3 mmHg or 2% of reading
Pulse: ± 5% of reading
Manual by inflation bulb
Automatic pressure release valve
Push button valve
Capacitive Type Pressure Transducer
4 alkaline batteries (type LR6 1.5 V)
Capacity of a new batteries is approx.
1000 readings
10°C to 40°C
30 to 85% RH
-20°C to 60°C
Approximately 170 g (including battery)
Approximately 12.5 (W) x 8.5 (D) x 6.0 (H) cm
For arms 22 to 33 cm circumference
Cuff, Type M (CR13); inflation system,
instruction manual
OMRON Matsusaka Co. Ltd., Japan
English