Your new blood pressure monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood movement through your brachial artery and converts the movements into a digital reading. An oscillometric monitor does not need a stethoscope so the monitor is simple to use.
This device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with arm circumference ranging from 9 inches to 17 inches (22 cm to 42 cm). The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.
Please read the Important Safety Information in this instruction manual before using the unit.
Please follow this instruction manual thoroughly for your safety.
Please keep for future reference. For specific information about your own blood pressure, CONSULT YOUR PHYSICIAN.
Contraindications
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property.
Data Transmission
Battery Handling and Usage
Contents:
Monitor, 4 "AAA" batteries, instruction manual, quick start guide
Monitor:
Display:
![]() | Connection symbol | ![]() | Irregular heartbeat symbol |
![]() | Connection OFF symbol | ![]() | Battery symbol (low) |
![]() | Connection ON symbol | ![]() | Battery symbol (depleted) |
![]() | Movement error symbol | ![]() | Deflation symbol |
![]() | SYNC symbol | ![]() | Heartbeat symbol (Flashes during measurement.) |
Connection symbol ()
Appears when transferring the data or pairing with your smart device.
Connection OFF symbol ()
Appears when Bluetooth is off.
Connection ON symbol ()
Appears when Bluetooth is on.
Movement error symbol ()
The movement error symbol will display when your body is moving during measurement. If the movement symbol displays, remove the monitor and wait for 2 to 3 minutes. After 2 to 3 minutes, remain still and take another measurement.
SYNC symbol ()
Irregular heartbeat symbol ()
When the monitor detects an irregular rhythm two or more times during the measurement, the irregular heartbeat symbol will appear on the display along with the measurement values. An irregular heartbeat rhythm is defined as a rhythm that is 25% less or 25% more than the average rhythm detected while the monitor is measuring the systolic and diastolic blood pressure. If the irregular heartbeat symbol displays with your measurement results, we recommend you consult your physician. Follow the directions of your physician.
Battery symbol (low) ()
Appears when the batteries are low.
Battery symbol (depleted) ()
Appears when the batteries are depleted.
Deflation symbol ()
Flashes while arm cuff is deflating.
Heartbeat symbol ()
Flashes while taking a measurement.
2013 ESH/ESC Guidelines for the management of arterial hypertension Definitions of hypertension by office and home blood pressure levels
Office | Home | |
Systolic Blood Pressure | ≥140 mmHg | ≥135 mmHg |
Diastolic Blood Pressure | ≥90 mmHg | ≥85 mmHg |
These are from statistical values for blood pressure.
To help ensure an accurate reading, follow these directions:
Avoid bathing, drinking alcohol or caffeine, smoking, exercising and eating for 30 minutes before taking a measurement.
Rest for at least 5 minutes before taking the measurement.
Remove tight-fitting or thick clothing from your arm while taking a measurement.
To take a measurement, you need to be relaxed and comfortably seated, at a comfortable room temperature.
You can use this blood pressure monitor without pairing to a smart device.
Notes:
DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat high blood pressure.
Always consult your physician. Self-diagnosis of measurement results and self-treatment are dangerous.
The date and time on your monitor will automatically be set when you pair it with your smart device.
Review the list of compatible smart devices at: OmronHealthcare.com
Notes:
DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat high blood pressure.
Always consult your physician. Self-diagnosis of measurement results and self-treatment are dangerous.
Data Transfer With Multiple Smart Devices
Once you transfer your data to a smart device, that data can no longer be transferred again to the same smart device, nor to another smart device.
About Multiple Users
To share your blood pressure monitor with other people, please refer to the "Help" section in the "OMRON connect" app.
Turn off the Bluetooth® in your monitor in the following areas where use of wireless equipment is prohibited.
Your monitor comes with the Bluetooth® already turned on.
To delete all the information stored in your monitor, follow the instructions below. Make sure that your monitor is turned off. It will be off when the screen is not showing numbers or symbols.
Error Display | Cause | Solution |
![]() | Irregular heartbeat is detected. | Remove the arm cuff. Wait 2 - 3 minutes and then take another measurement. Repeat the steps in section 2.2. If this error continues to appear, contact your physician. |
![]() | Movement during measurement. | Carefully read and repeat the steps in section 4.2. |
![]() | The batteries are low. | Recommend replacement of all 4 batteries with new ones. |
![]() | The batteries are depleted or battery polarities are not properly aligned. | Immediately replace 4 batteries with new ones. |
Confirm the batteries are properly inserted with polarities correctly aligned. | ||
![]() | Connection failure. Data is not being transmitted. | Refer to "Connection failure. Data is not being transmitted." |
![]() | Arm cuff is applied too loosely. | Apply the arm cuff tighter. |
![]() | The arm cuff was inflated exceeding the maximum allowable pressure, and then deflated automatically. | Do not touch the arm cuff while taking a measurement. |
![]() | Movement during measurement. | Repeat measurement. Remain still and do not talk during measurement. |
![]() | Clothing is interfering with the arm cuff. | Remove any clothing interfering with the arm cuff. |
![]() *: "XX" will change to display an error number, 6 and higher. | Device error. | Please contact customer service. |
Problem | Cause and Solution |
No power. No display appears on the monitor. | Replace all batteries with new ones. Check the battery installation for proper placement of the battery polarities. |
Measurement values appear too high or too low. | Blood pressure varies constantly. Many factors including stress, time of day, and how you apply the arm cuff, may affect your blood pressure. |
Connection failure. / Data is not being transmitted. | The blood pressure monitor might not be properly placed within the smart device's transmission range and is too far from the smart device. If there are no causes of data transmission interference found near the blood pressure monitor, move the blood pressure monitor within 16 ft. (5 m) of the smart device and try again. |
The Bluetooth® feature of the smart device is turned off. Turn on the Bluetooth® feature of the smart device and your monitor, and try sending the data again. | |
The Bluetooth® feature of the blood pressure monitor is turned off. Refer to "To turn "ON" the Bluetooth®" in "Turning "OFF" the Bluetooth®". | |
The blood pressure monitor did not pair successfully to the smart device. Try to pair the devices again. Refer to "Pair Your Monitor with a Smart Device". Confirm your smart device compatibility with this monitor at OmronHealthcare.com | |
The application on the smart device is not ready. Check the application then try sending the data again. Refer to "Pair Your Monitor with a Smart Device". If the "Error" symbol still displays after checking the application, contact customer service. |
To protect your monitor from damage, please follow the directions below:
Do not disassemble or attempt to repair the monitor or components.
Do not store the monitor in the following situations:
Dispose of the monitor and components according to applicable local regulations. Unlawful disposal may cause environmental pollution.
Model | BP7000 ![]() |
Display | OLED display |
Cuff pressure range | Pressure: 0 to 299 mmHg |
Measurement range | Pressure: 40 to 260 mmHg Pulse: 40 to 180 beats / min. |
Accuracy | Pressure: ±3 mmHg Pulse: ±5% of display reading |
Inflation | Fuzzy-logic controlled by electric pump |
Deflation | Automatic rapid deflation |
Measurement method | Oscillometric method |
Transmission method | Bluetooth ® Smart |
Wireless communication | Frequency range: 2.4 GHz (2400 - 2483.5 MHz) Modulation: GFSK Effective radiated power: <20 dBm |
IP classification | IP 22 |
Power source | 4 "AAA" batteries 1.5V |
Battery life | Approximately 300 measurements (using new alkaline batteries) |
Durable period (Service life) Monitor | 5 years |
Operating conditions | 50°F to 104°F (10°C to 40°C) / 15 to 90% RH (non-condensing) / 800 to 1060 hPa |
Storage / Transport conditions | -4°F to 140°F (-20°C to 60°C) / 10 to 90% RH (non-condensing) |
Weight | Monitor: Approximately 8 1/2 oz. (240 g) not including batteries |
Dimensions | Monitor: Approximately 3 1/2" (w) ![]() ![]() ![]() ![]() |
Arm circumference | 9" to 17" (22 to 42 cm) |
Contents | Monitor, 4 "AAA" batteries, instruction manual, quick start guide |
Applied part | Type BF |
Protection against electric shock | Internally powered ME equipment |
Notes:
About a wireless communication interference
This Product operates in the unlicensed ISM band at 2.4GHz. In case this Product is used around the other wireless devices including microwave and wireless LAN, which operate same frequency band of this Product, there is a possibility that interference occurs between this Product and such other devices. If such interference occurs, please stop the operation of other devices or relocate this Product before using this Product or do not use this Product around the other wireless devices.
![]() | Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. |
![]() | Indicates a potentially hazardous situation, which if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property. |
Note | Provides additional information to clarify a point in the text. |
![]() | Please consult this instruction manual |
![]() | Applied part - Type BF Degree of protection against electric shock (leakage current) |
![]() | To indicate generally elevated, potentially hazardous, levels of non-ionizing radiation, or to indicate equipment or systems e.g. in the medical electrical area that include RF transmitters or that intentionally apply RF electromagnetic energy for diagnosis or treatment. |
![]() | Catalogue number |
![]() | Manufacturer |
![]() | Temperature limitation |
![]() | Humidity limitation |
![]() | Atmospheric pressure limitation |
![]() | Artery mark |
FCC CAUTION
Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.
Note:
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
This transmitter must not be co-located or operated in conjunction with any other antenna or transmitter.
This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment and meets the FCC radio frequency (RF) Exposure Guidelines. This equipment has very low levels of RF energy that are deemed to comply without testing of specific absorption ratio (SAR).
Your BP7000 Upper Arm Blood Pressure Monitor, excluding batteries, is warranted to be free from defects in materials and workmanship appearing within 5 years from the date of purchase, when used in accordance with the instructions provided with the monitor. The above warranty extends only to the original retail purchaser.
We will, at our option, replace without charge any monitor covered by the above warranty. Replacement is our only responsibility and your only remedy under the above warranty.
To obtain warranty service contact Customer Service by calling 1-800-634-4350 for the address of the inspection center and the return shipping and handling fee.
Enclose the original printed receipt. Include a letter, with your name, address, phone number, and description of the specific problem. Pack the product carefully to prevent damage in transit. Because of possible loss in transit, we recommend insuring the product with return receipt requested.
THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS THAT MAY BE IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION TO THE PERIOD OF THE ABOVE EXPRESS WARRANTY. OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER SPECIAL, INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS, EXPENSES OR DAMAGES.
This warranty provides you with specific legal rights, and you may have other rights that vary by jurisdiction. Because of special local requirements, some of the above limitations and exclusions may not apply to you.
FOR CUSTOMER SERVICE | ||
Visit our web site at: | OmronHealthcare.com | |
Call toll free: | 1-800-634-4350 |
OMRON Blood Pressure Monitor (BPM)
Information for accompanying documents in the scope of IEC60601-1-2:2007
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC's and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation. Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to prevent unsafe product situations, the IEC60601-1-2 standard has been implemented. This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices. Medical devices manufactured by OMRON Healthcare conform to this IEC60601-1-2:2007 standard for both immunity and emissions. Nevertheless, special precautions need to be observed:
The BPM may be interfered with by other equipment, even if that other equipment complies with CISPR EMISSION requirements. | |||||||||||||
Guidance and manufacturer's declaration - electromagnetic emissions | |||||||||||||
OMRON battery-operated BPM is intended for use in the electromagnetic environment specified below. The customer or the user of this OMRON battery-operated BPM should assure that it is used in such environment. | |||||||||||||
Emissions test | Compliance | Electromagnetic environment - guidance | |||||||||||
RF emissions CISPR 11 | Group 1 | The OMRON battery-operated BPM uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. | |||||||||||
RF emissions CISPR 11 | Class B | The OMRON battery-operated BPM is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. | |||||||||||
Harmonic emissions IEC 61000-3-2 | Not Applicable. | ||||||||||||
Voltage fluctuations/ flicker emissions IEC61000-3-3 | Not Applicable. | ||||||||||||
Guidance and manufacturer's declaration - electromagnetic immunity | |||||||||||||
OMRON battery-operated BPM is intended for use in the electromagnetic environment specified below. The customer or the user of this OMRON battery-operated BPM should assure that it is used in such environment. | |||||||||||||
Immunity test | IEC 60601 test level | Compliance level | Electromagnetic environment - guidance | ||||||||||
Electrostatic discharge (ESD) IEC 61000-4-2 | Contact discharge: ±6 kV Air discharge: ±8 kV | Contact discharge: ±8 kV Air discharge: ±15 kV | The relative humidity should be at least 5%. | ||||||||||
Electrical fast transient/burst IEC 61000-4-4 | ±2 kV for power supply lines ±1 kV for input/output lines | Not Applicable. | Not Applicable. | ||||||||||
Surge IEC 61000-4-5 | ±1 kV line(s) to line(s) ±2 kV line(s) to earth | Not Applicable. | Not Applicable. | ||||||||||
Voltage dips, short interruptions and voltage variations on power supply inputlines IEC 61000-4-11 | <5% UT (>95% dip in UT) for 0.5 cycle | Not Applicable. | Not Applicable. | ||||||||||
40% UT (60% dip in UT) for 5 cycles | Not Applicable. | ||||||||||||
70% UT (30% dip in UT) for 25 cycles | Not Applicable. | ||||||||||||
<5% UT (>95% dip in UT) for 5 sec. | Not Applicable. | ||||||||||||
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 | 3 A/m | 30 A/m | Magnetic fields from common appliances are not expected to affect the device. | ||||||||||
Note: UT is the A.C. mains voltage prior to application of the test level. | |||||||||||||
Guidance and manufacturer's declaration - electromagnetic immunity | |||||||||||||
OMRON battery-operated BPM is intended for use in the electromagnetic environment specified below. The customer or the user of this OMRON battery-operated BPM should assure that it is used in such environment. | |||||||||||||
Immunity test | IEC 60601 test level | Compliance level | Electromagnetic environment - guidance | ||||||||||
Conducted RF IEC 61000-4-6 Radiated RF IEC 61000-4-3 | 3 V rms 150 kHz to 80 MHz 3 V/m | Not Applicable. 10 V/m | Portable and mobile RF communications equipment should be used no closer to any part of the OMRON battery-operated BPM including cables, than the recommended separation distance calculated from the equation appropriate to the frequency of the transmitter. Recommend separation distance Not Applicable. d = 0.35 80 MHz to 800 MHz d = 0.70 800 MHz to 2.5 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol: | ||||||||||
Note1: At 80 MHz and 800 MHz, the higher frequency range applies. Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people. | |||||||||||||
a Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones and land mobile radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the OMRON battery-operated BPM is used exceeds the applicable RF compliance level above, the OMRON battery-operated BPM should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the OMRON battery-operated BPM. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 6 V/m. | |||||||||||||
Recommended separation distance between portable and mobile RF communications equipment and the OMRON battery-operated BPM | |||||||||||||
OMRON battery-operated BPM is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of this OMRON battery-operated BPM can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the OMRON battery-operated BPM as recommended below, according to the maximum output power of the communications equipment. | |||||||||||||
Output Power of Transmitter in Watt | Separation distance according to frequency of transmitter in meter | ||||||||||||
150 kHz to 80 MHz Not Applicable. | 80 MHz to 800 MHz d = 0.35 ![]() | 800 MHz to 2.5 GHz d = 0.70 ![]() | |||||||||||
0.01 | Not Applicable. | 0.035 | 0.070 | ||||||||||
0.1 | Not Applicable. | 0.11 | 0.22 | ||||||||||
1 | Not Applicable. | 0.35 | 0.70 | ||||||||||
10 | Not Applicable. | 1.1 | 2.2 | ||||||||||
100 | Not Applicable. | 3.5 | 7.0 | ||||||||||
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people. |
The Bluetooth® Smart word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by OMRON HEALTHCARE Co., Ltd. is under license. Other trademarks and trade names are those of their respective owners.
Apple and the Apple logo are trademarks of Apple Inc., registered in the U.S. and other countries. App Store is a service mark of Apple Inc.
Android and the Google Play logo are trademarks of Google Inc.
Manufactured by | OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN |
Distributed by | OMRON HEALTHCARE, INC. 1925 West Field Court, Lake Forest, IL 60045 U.S.A. 1-800-634-4350 OmronHealthcare.com |