Omron M3 Automatic Blood Pressure Monitor Manual

Omron M3 Automatic Blood Pressure Monitor Manual

Also See for M3:

Advertisement

Omron M3 Automatic Blood Pressure Monitor Manual

Introduction

Thank you for purchasing the OMRON M3 Automatic Blood Pressure Monitor.
The OMRON M3 is a compact, fully automatic blood pressure monitor, operating on the oscillometric principle. It measures your blood pressure and pulse rate simply and quickly. For comfortable controlled inflation without the need of pressure pre-setting or re-inflation the device uses its advanced "IntelliSense" technology.

Intended Use

This device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population who can understand this instruction manual with the arm circumference range printed on the arm cuff. The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with the measurement result.

information Please read this instruction manual thoroughly before using the device.
Please keep for future reference. For specific information about your own blood pressure, CONSULT YOUR PHYSICIAN.

Important Safety Information


Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.

General Usage

warning Consult your physician before using the device in pregnancy including pre-eclampsia, or if diagnosed with arrhythmia or arteriosclerosis.
warning Do not use the device on the injured arm or the arm under medical treatment. Do not apply the arm cuff on the arm while being on an intravenous drip or blood transfusion.
warning Consult your physician before using the device on the arm with an arterio-venous (A-V) shunt.
warning Do not use the device with other medical electrical (ME) equipment simultaneously.
warning Do not use the device in the area the HF surgical equipment, MRI, or CT scanner exists, or in the oxygen rich environment.
warning The air tube or the AC adapter cable may cause accidental strangulation in infants.
warning Contained small parts that may cause a choking hazard if swallowed by infants.
warning Always consult your physician. Self-diagnosis of measurement results and self-treatment are dangerous.
warning People with severe blood flow problems, or blood disorders, should consult a physician before using the device, as the arm cuff inflation can cause bruising.
warning If there are any abnormalities during the measurement, remove the arm cuff.
warning Do not use this device on infants or persons who cannot express their intentions.
warning Do not inflate the arm cuff more than necessary.
warning Do not use the device for any purpose other than measuring blood pressure. Use only the approved arm cuff for this device. Use of other arm cuffs may result in incorrect measurement results.
warning Do not use a mobile phone or other devices that emit electromagnetic fields, near the device. This may result in incorrect operation of the device.
warning Do not disassemble the monitor or arm cuff.
warning Do not use in a location with moisture, or a location where water may splash on the device. This may damage the device.
warning Do not use the device in a moving vehicle (car, airplane).
warning Do not take measurements more than necessary. It may cause bruising due to blood flow interference.
warning Consult your physician before using the device if you had a mastectomy.

AC Adapter (optional) Usage

warning Do not use the AC adapter if the device or the power cord is damaged. Turn off the power and unplug the power cord immediately.
warning Plug the AC adapter into the appropriate voltage outlet. Do not use in a multi-outlet plug.
warning Never plug in or unplug the power cord from the electric outlet with wet hands.


Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property.

warning Fully insert the power plug into the outlet.
warning When disconnecting the power plug from the outlet, do not pull the power cord.
warning Be sure to pull from the power plug safely.
warning When handling the power cord, take care not to do the following: Do not damage. Do not break it.

Do not damage.
Do not tamper with it.
Do not forcibly bend or pull.

Do not twist.
Do not pinch.

Do not break it.
Do not bundle during use.
Do not place under heavy objects.

warning Wipe the dust off from the power plug.
warning Unplug monitor when not in use.
warning Disconnect the power plug before cleaning.
warning Use only the original adapter designed for this device. Use of unsupported adapters may damage and/or may be hazardous to the device.

Battery Usage

warning Do not insert the batteries with their polarities incorrectly aligned.

warning Use only 4 "AA" alkaline or manganese batteries with this device. Do not use other types of batteries. Do not use new and used batteries together.

warning Remove the batteries if the device will not be used for three months or more.

General Precautions

  • Do not forcibly crease the arm cuff or the air tube excessively.
  • Do not press the air tube while taking a measurement.
  • To unplug the air plug, pull on the air plug at the connection with the monitor, not the tube itself.
  • Do not drop the monitor or subject device to strong shocks or vibrations.
  • Do not inflate the arm cuff when it is not wrapped around your arm.
  • Do not use the device outside the specified environment. It may cause an inaccurate result.
  • Read and follow the "Important information regarding Electro Magnetic Compatibility (EMC)" in the "Specifications".
  • Read and follow the "Correct Disposal of This Product" in "Specifications" when disposing of the device and any used accessories or optional parts.

Know Your Device

Contents

Monitor, arm cuff, instruction manual, storage case, battery set, blood pressure pass

Monitor

Know Your Device - Monitor

  1. Display
  2. Memory button
  3. START/STOP button
  4. Date/Time setting button
  5. USER ID selection switch
  6. Cuff wrap guide lamp
  7. Up/Down buttons
  8. Blood pressure colour indicator
  9. Battery compartment
  10. AC adapter jack (for optional AC adapter)
  11. Air jack

Arm cuff

Know Your Device - Arm Cuff

  1. Arm cuff (Arm circumference 22 - 42 cm)
  2. Air plug
  3. Air tube

Display

Know Your Device - Display

  1. USER ID symbol
  2. Systolic blood pressure
  3. Diastolic blood pressure
  4. Low battery symbol
  5. Cuff wrap guide
  6. Heartbeat symbol
    (Flashes during measurement.)
  7. Memory symbol
  8. Average value symbol
  9. Date/Time display
  10. Movement error symbol
  11. Irregular heartbeat symbol
  12. Blood pressure level indicator
  13. Deflation symbol
  14. Pulse display / Memory number

Display symbols

Irregular Heartbeat Symbol ( )
When the monitor detects an irregular rhythm two or more times during the measurement, the irregular heartbeat symbol ( ) will appear on the display with the measurement values.
An irregular heartbeat rhythm is defined as a rhythm that is 25% less or 25% more than the average rhythm detected while the monitor is measuring the systolic and diastolic blood pressure.
If the irregular heartbeat symbol ( ) displays with your measurement results, we recommend you consult your physician. Follow the directions of your physician.
Heartbeat Diagram

Movement Error Symbol ( )
The movement error symbol is displayed if you move your body during the measurement. Please remove the arm cuff, and wait 2 - 3 minutes. Take another measurement, remain still during measurement.

Average Value Symbol ( )
The average value symbol is displayed when you press and hold the memory button for more than 3 seconds. The most recent average value appears on the display screen.

Cuff Wrap Guide Lamp ( )
If the cuff was wrapped too loosely, it may cause unreliable values. If the wrapping of cuff is too loose, the cuff wrap guide lamp does not light.
Otherwise lights in "green". This is the function which is used as an aid in determining if the cuff is wrapped snugly enough.

Blood Pressure Colour Indicator
If your systolic or diastolic pressure is above the standard range (135 mmHg for the systolic blood pressure and/or 85 mmHg for the diastolic blood pressure), the blood pressure colour indicator will light in "orange" when the measurement result is displayed. If they are within the standard range, no light will appear.

The JNC7* Guideline recommends the following guideline.

General Guidelines for Blood Pressure
Prehypertension at Office Hypertension at Home
Systolic Blood Pressure 120 - 139 mmHg 135 mmHg
Diastolic Blood Pressure 80 - 89 mmHg 85 mmHg

These are form statistical values for blood pressure.

* JNC7: The Seventh Report, 2003 Dec, of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of Hi Blood Pressure.

Preparation

Battery Installation

  1. Remove the battery cover.
  2. Insert 4 "AA" batteries as indicated in the battery compartment.
    Battery Installation
  3. Replace the battery cover.

Notes:

  • When the low battery symbol ( ) appears on the display, turn the monitor off, then replace all batteries at the same time. Long life alkaline batteries are recommended.
  • The measurement values continue to be stored in memory even after the batteries are replaced.
  • The supplied batteries may have a shorter life.

warning Disposal of used batteries should be carried out in accordance with the national/local regulations for the disposal of batteries.

Setting the Date and Time

Set the monitor to the correct date and time before taking a measurement for the first time.

  1. Press the button.
  2. Push or to change the year
    Push button to confirm the year and then the month flashes. Repeat the same steps to change the month, day, hour, and minutes.
    Setting the Date and Time
  3. Press the START/STOP button to turn the monitor off.

Notes:

  • If the batteries have been replaced, the date and time setting will need to be reset.
  • If the date and time are not set, "-:--" appears during or after measurement.

Using the Device

Applying the Arm Cuff

Remove tight-fitting clothing or tight rolled up sleeve from your left upper arm.
Do not place the arm cuff over thick clothes.

  1. Insert the air plug into the air jack securely.
    Applying the Arm Cuff - Step 1
  2. Wrap the arm cuff firmly in place around your left upper arm.

    The bottom edge of the arm cuff should be 1 to 2 cm above the elbow.
    Air tube is on the inside of your arm and aligned with your middle finger.
    Applying the Arm Cuff - Step 2
  3. Secure closed with the fabric fastener.
    Applying the Arm Cuff - Step 3 - Secure Fastening

Notes:

  • When you take a measurement on the right arm, the air tube will be at the side of your elbow. Be careful not to rest your arm on the air tube.
  • The blood pressure can differ between the right arm and the left arm, and therefore also the measured blood pressure values can be different. OMRON recommends to always use the same arm for measurement. If the values between both arms differ substantially, please check with your physician which arm to use for your measurement.

How to Sit Correctly

To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature. Avoid bathing, drinking alcohol or caffeine, smoking, exercising or eating 30 minutes before taking a measurement.

  • Sit on a chair with your feet flat on the floor.
  • Sit upright with your back straight.
  • Sit with your back and arm being supported.
  • The arm cuff should be placed on your arm at the same level as your heart.
    How to Sit Correctly

Taking a Measurement

Notes:

  • To cancel a measurement, press the START/STOP button to release the air in the arm cuff.
  • Remain still while taking a measurement.

The monitor is designed to take measurements and store the measurement values in the memory for 2 people using USER ID 1 and USER ID 2.

  1. Select your User ID (1 or 2).
    Taking a Measurement - Step 1 - Select the User ID
  2. Press the START/STOP button.
    The arm cuff will start to inflate automatically.
    Taking a Measurement - Step 2 - Press START/STOP
    If your systolic pressure is more than 210 mmHg
    After the arm cuff starts to inflate, press and hold the START/STOP button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure.

Notes:

  • The monitor will not inflate above 299 mmHg.
  • Do not apply more pressure than necessary.
  1. Remove the arm cuff.
  2. Press the START/STOP button to turn the monitor off.
    The monitor automatically stores the measurement result in its memory.
    It will automatically turn off after 2 minutes.

Note: Wait 2-3 minutes before taking another measurement. Waiting between measurements allows the arteries to return to the condition prior to taking a measurement.

Using the Guest Mode

The monitor stores measurement values for 2 users in the memory. The guest mode can be used to take a single measurement for another user. No measurement values are stored in the memory when the guest mode is selected.

  1. Press and hold the START/STOP button for more than 3 seconds.
    The USER ID symbol and the Date/Time display will disappear.
  2. Release the START/STOP button when the Date/Time display turns off.
    The arm cuff will start to inflate automatically.

warning Always consult your physician. Self-diagnosis of measurement results and self-treatment are dangerous.

Using the Memory Function

The monitor automatically stores the results up to 60 sets for each user (1 and 2). It can also calculate an average value based on the last 3 measurement values taken within 10 minutes.

Notes:

  • If there are only 2 measurement values in the memory for that period, the average will be based on these 2 values.
  • If there is 1 measurement value in the memory for that period, this is displayed as the average.
  • If the memory is full, the monitor will delete the oldest value.
  • When viewing the measurement value taken without setting the date and time, "-:--" is displayed instead of the date and time.

To View the Measurement Values Stored in Memory

  1. Select your User ID (1 or 2).
  2. Press the button.
    The Memory number appears for a second before the pulse rate is displayed.
    The newest set is numbered "1".
    Viewing the Measurement Values Stored in Memory

Note: The cuff wrap guide result appears on the display with the measurement values. The cuff wrap guide lamp will not light.
: The arm cuff was wrapped snugly enough.
: The arm cuff was loose or not wrapped correctly.

  1. Press the or button to view the values stored in memory.
    : To the older values
    : To the more recent values

To View the Average Value

  1. Select your User ID (1 or 2).
  2. Press and hold the button for more than 3 seconds.

Notes:

  • If the previous measurement was taken without setting the date and time, the average value is not calculated.
  • If there are no measurements results stored in the memory, the screen to the right is displayed.

To Delete All the Values Stored in Memory

  1. Select your User ID (1 or 2).
  2. Press the Memory button, while the memory symbol ( ) appears.
  3. While holding the button down, press the START/STOP button for more than 3 seconds.
    Deleting All the Values Stored in Memory

Note: You cannot partially delete the values stored in the memory.
All values for the user you select will be deleted.

Error Messages

Error Display Cause Solution
Irregular heartbeats are detected. Remove the arm cuff. Wait 2 - 3 minutes and then take another measurement. Repeat the steps in "Taking a Measurement" section. If this error continues to appear, contact your physician.
Movement during measurement. Carefully read and repeat the steps in "Taking a Measurement" section.
The batteries are low. You should replace the batteries with new ones ahead of time.
The batteries are exhausted. You should replace the batteries with new ones at once.
Air plug disconnected. Insert the plug securely.
Arm cuff is applied too loosely. Apply the arm cuff tighter.
Air is leaking from the arm cuff. Replace the cuff with a new one.
Movement during measurement
and the arm cuff has not been inflated sufficiently.
Repeat measurement. Remain still and do not talk during measurement.
If "E2" appears repeatedly, inflate the cuff manually until it is 30 to 40 mmHg above your previous measurement result.
The arm cuff was inflated above 299 mmHg when inflating the cuff manually. Do not inflate the cuff above 299 mmHg.
Movement during measurement. Repeat measurement. Remain still and do not talk during measurement.
Clothing is interfering with the arm cuff. Remove any clothing interfering with the arm cuff.
Device error. Contact your OMRON retail outlet or distributor.

Troubleshooting

Problem Cause Solution

The measurement result is extremely high (or low)

Arm cuff is applied too loosely. Apply the arm cuff tighter.
Movement or talking during measurement. Remain still and do not talk during measurement.
Clothing is interfering with the arm cuff. Remove any clothing interfering with the arm cuff.

Arm cuff pressure does not rise

The air connector is not securely connected into the air jack. Make sure that the air tube is connected securely.
Air is leaking from the arm cuff. Replace the arm cuff with a new one.

The cuff wrapping guide lamp does not light

Arm cuff deflates too soon

The arm cuff is loose. Apply the cuff correctly so that it is firmly wrapped around the arm.
Cannot measure or the results are too low or too high. The arm cuff has not been inflated sufficiently. Inflate the cuff so that it is 30 to 40 mmHg above your previous measurement result.

Nothing happens
when you press the buttons

The batteries are empty. Replace the batteries with new ones.
The batteries have been inserted incorrectly. Insert the batteries with the correct (+/-) polarity.

Other problems

  • Press the START/STOP button and repeat measurement.
  • Replace the batteries with new ones. If the problem continues, contact your OMRON retail outlet or distributor.

Maintenance and Storage

Maintenance

To protect your device from damage, please observe the following:

  • Store the device and the components in a clean, safe location.

  • Do not use any abrasive or volatile cleaners.
  • Do not wash the device and any components or immerse them in water.
  • Do not use petrol, thinners or similar solvents to clean the device.
  • Use a soft and dry cloth, or a soft and moistened cloth and neutral soap to clean on the monitor and the arm cuff.
  • Changes or modification not approved by the manufacturer will void the user warranty. Do not disassemble or attempt to repair the device or components. Consult your authorised OMRON retail outlet or distributor.

Calibration and Service

  • The accuracy of this blood pressure monitor has been carefully tested and is designed for a long service life.
  • It is generally recommended to have the device inspected every 2 years to ensure correct functioning and accuracy. Please consult your authorised OMRON retail outlet or distributor.

Storage

Keep the device in its storage case when not in use.

  1. Unplug the air plug from the air jack.
  2. Gently fold the air tube into the arm cuff.

    Note: Do not bend or crease the air tube excessively.
  3. Place the monitor and the arm cuff in the storage case.

Do not store the device in the following situations:

  • If the device is wet.
  • Locations exposed to extreme temperatures, humidity, direct sunlight, dust or corrosive vapours.
  • Locations exposed to vibrations, shocks or where it will be at an angle.

Optional Medical Accessories

(within the scope of EC Medical Device Directive 93/42/EEC)

Optional Medical Accessories

Using the Optional AC Adapter

  1. Insert the AC adapter plug into the AC adapter jack on the rear side of the monitor.
  2. Plug the AC adapter into an electrical outlet.
    Using the Optional AC Adapter

To disconnect the AC adapter, unplug the AC adapter from the electrical outlet first and then remove the AC adapter plug from the monitor.

Specifications

Product description Automatic Blood Pressure Monitor
Model OMRON M3 (HEM-7131-E)
Display LCD Digital Display
Measurement method Oscillometric method
Measurement range Pressure: 0 to 299 mmHg
Pulse: 40 to 180 beats/min.
Accuracy Pressure: ±3 mmHg
Pulse: ±5% of display reading
Inflation Fuzzy-logic controlled by electric pump
Deflation Automatic pressure release valve
Memory 60 measurements with date and time for each user (1 and 2)
Rating DC6V 4W
Power source 4 "AA" batteries 1.5V or optional AC adapter
(Adapter S-9515336-9, INPUT AC100-240V 50/60Hz 0.12A)
(Adapter UK-9983666-5, INPUT AC100-240V 50/60Hz 15VA)
Battery life Approx. 1000 measurements (using new alkaline batteries)
Applied part = Type BF
Protection against electric shock Internally powered ME equipment (When using only the batteries)
= Class II ME equipment (Optional AC adapter)
Operating temperature/ humidity +10 to +40°C / 30 to 85% RH
Storage temperature/ humidity/ air pressure -20 to +60°C / 10 to 95% RH / 700 to 1060 hPa
IP classification IP 20
Weight Monitor: Approx. 280 g without batteries
Arm cuff: Approx. 170 g
Outer dimensions Monitor: Approx. 107 (w) mm x 79 (h) mm x 141 (l) mm
Arm cuff: Approx. 145 mm x 594 mm
Cuff circumference 22 to 42 cm
Cuff/ Tube material Nylon, polyester, polyvinyl chloride
Package contents Monitor, arm cuff, instruction manual, storage case, battery set, blood pressure pass

Notes:

  • These specifications are subject to change without notice.
  • In the clinical validation study, the 5th phase was used on 85 subjects for determination of diastolic blood pressure.
  • This device has not been validated for use on pregnant patients.

  • This device fulfils the provisions of EC directive 93/42/EEC (Medical Device Directive).
  • This blood pressure monitor is designed according to the European Standard EN1060, Non-invasive sphygmomanometers Part 1: General Requirements and Part 3: Supplementary requirements for electromechanical blood pressure measuring systems.
  • This OMRON product is produced under the strict quality system of OMRON HEALTHCARE Co. Ltd., Japan. The Core component for OMRON blood pressure monitors, which is the Pressure Sensor, is produced in Japan.

Important information regarding Electro Magnetic Compatibility (EMC)

With the increased number of electronic devices such as PC's and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation. Medical devices should also not interfere with other devices.

In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to prevent unsafe product situations, the EN60601-1-2:2007 standard has been implemented. This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices.

This medical device manufactured by OMRON HEALTHCARE conforms to this EN60601-1-2:2007 standard for both immunity and emissions. Nevertheless, special precautions need to be observed:

  • Do not use mobile (cellular) telephones and other devices, which generate strong electrical or electromagnetic fields, near the medical device. This may result in incorrect operation of the device and create a potentially unsafe situation. Recommendation is to keep a minimum distance of 7 m. Verify correct operation of the device in case the distance is shorter.

Further documentation in accordance with EN60601-1-2:2007 is available at OMRON HEALTHCARE EUROPE at the address mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.

Correct Disposal of This Product (Waste Electrical & Electronic Equipment)

This marking shown on the product or its literature, indicates that it should not be disposed of, with other household wastes at the end of its working life. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this product from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources.

Household users should contact either the retailer where they purchased this product, or their local government office, for details of where and how they can return this item for environmentally safe recycling.

Business users should contact their supplier and check the terms and conditions of the purchase contract. This product should not be mixed with other commercial wastes for disposal.

Documents / Resources

References

Download manual

Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.