Thank you for purchasing the OMRON M3 Automatic Blood Pressure Monitor.
The OMRON M3 is a compact, fully automatic blood pressure monitor, operating on the oscillometric principle. It measures your blood pressure and pulse rate simply and quickly. For comfortable controlled inflation without the need of pressure pre-setting or re-inflation the device uses its advanced "IntelliSense" technology.
This device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population who can understand this instruction manual with the arm circumference range printed on the arm cuff. The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with the measurement result.
Please read this instruction manual thoroughly before using the device.
Please keep for future reference. For specific information about your own blood pressure, CONSULT YOUR PHYSICIAN.
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Consult your physician before using the device in pregnancy including pre-eclampsia, or if diagnosed with arrhythmia or arteriosclerosis.
Do not use the device on the injured arm or the arm under medical treatment. Do not apply the arm cuff on the arm while being on an intravenous drip or blood transfusion.
Consult your physician before using the device on the arm with an arterio-venous (A-V) shunt.
Do not use the device with other medical electrical (ME) equipment simultaneously.
Do not use the device in the area the HF surgical equipment, MRI, or CT scanner exists, or in the oxygen rich environment.
The air tube or the AC adapter cable may cause accidental strangulation in infants.
Contained small parts that may cause a choking hazard if swallowed by infants.
Always consult your physician. Self-diagnosis of measurement results and self-treatment are dangerous.
People with severe blood flow problems, or blood disorders, should consult a physician before using the device, as the arm cuff inflation can cause bruising.
If there are any abnormalities during the measurement, remove the arm cuff.
Do not use this device on infants or persons who cannot express their intentions.
Do not inflate the arm cuff more than necessary.
Do not use the device for any purpose other than measuring blood pressure. Use only the approved arm cuff for this device. Use of other arm cuffs may result in incorrect measurement results.
Do not use a mobile phone or other devices that emit electromagnetic fields, near the device. This may result in incorrect operation of the device.
Do not disassemble the monitor or arm cuff.
Do not use in a location with moisture, or a location where water may splash on the device. This may damage the device.
Do not use the device in a moving vehicle (car, airplane).
Do not take measurements more than necessary. It may cause bruising due to blood flow interference.
Consult your physician before using the device if you had a mastectomy.
Do not use the AC adapter if the device or the power cord is damaged. Turn off the power and unplug the power cord immediately.
Plug the AC adapter into the appropriate voltage outlet. Do not use in a multi-outlet plug.
Never plug in or unplug the power cord from the electric outlet with wet hands.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property.
Fully insert the power plug into the outlet.
When disconnecting the power plug from the outlet, do not pull the power cord.
Be sure to pull from the power plug safely.
When handling the power cord, take care not to do the following: Do not damage. Do not break it.
Do not damage.
Do not tamper with it.
Do not forcibly bend or pull.
Do not twist.
Do not pinch.
Do not break it.
Do not bundle during use.
Do not place under heavy objects.
Wipe the dust off from the power plug.
Unplug monitor when not in use.
Disconnect the power plug before cleaning.
Use only the original adapter designed for this device. Use of unsupported adapters may damage and/or may be hazardous to the device.
Do not insert the batteries with their polarities incorrectly aligned.
Use only 4 "AA" alkaline or manganese batteries with this device. Do not use other types of batteries. Do not use new and used batteries together.
Remove the batteries if the device will not be used for three months or more.
Monitor, arm cuff, instruction manual, storage case, battery set, blood pressure pass
Irregular Heartbeat Symbol ( )
When the monitor detects an irregular rhythm two or more times during the measurement, the irregular heartbeat symbol ( ) will appear on the display with the measurement values.
An irregular heartbeat rhythm is defined as a rhythm that is 25% less or 25% more than the average rhythm detected while the monitor is measuring the systolic and diastolic blood pressure.
If the irregular heartbeat symbol ( ) displays with your measurement results, we recommend you consult your physician. Follow the directions of your physician.
Movement Error Symbol ( )
The movement error symbol is displayed if you move your body during the measurement. Please remove the arm cuff, and wait 2 - 3 minutes. Take another measurement, remain still during measurement.
Average Value Symbol ( )
The average value symbol is displayed when you press and hold the memory button for more than 3 seconds. The most recent average value appears on the display screen.
Cuff Wrap Guide Lamp ( )
If the cuff was wrapped too loosely, it may cause unreliable values. If the wrapping of cuff is too loose, the cuff wrap guide lamp does not light.
Otherwise lights in "green". This is the function which is used as an aid in determining if the cuff is wrapped snugly enough.
Blood Pressure Colour Indicator
If your systolic or diastolic pressure is above the standard range (135 mmHg for the systolic blood pressure and/or 85 mmHg for the diastolic blood pressure), the blood pressure colour indicator will light in "orange" when the measurement result is displayed. If they are within the standard range, no light will appear.
The JNC7* Guideline recommends the following guideline.
General Guidelines for Blood Pressure | ||
Prehypertension at Office | Hypertension at Home | |
Systolic Blood Pressure | 120 - 139 mmHg | 135 mmHg |
Diastolic Blood Pressure | 80 - 89 mmHg | 85 mmHg |
These are form statistical values for blood pressure.
* JNC7: The Seventh Report, 2003 Dec, of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of Hi Blood Pressure.
Notes:
Disposal of used batteries should be carried out in accordance with the national/local regulations for the disposal of batteries.
Set the monitor to the correct date and time before taking a measurement for the first time.
Notes:
Remove tight-fitting clothing or tight rolled up sleeve from your left upper arm.
Do not place the arm cuff over thick clothes.
Notes:
To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature. Avoid bathing, drinking alcohol or caffeine, smoking, exercising or eating 30 minutes before taking a measurement.
Notes:
The monitor is designed to take measurements and store the measurement values in the memory for 2 people using USER ID 1 and USER ID 2.
Notes:
Note: Wait 2-3 minutes before taking another measurement. Waiting between measurements allows the arteries to return to the condition prior to taking a measurement.
The monitor stores measurement values for 2 users in the memory. The guest mode can be used to take a single measurement for another user. No measurement values are stored in the memory when the guest mode is selected.
Always consult your physician. Self-diagnosis of measurement results and self-treatment are dangerous.
The monitor automatically stores the results up to 60 sets for each user (1 and 2). It can also calculate an average value based on the last 3 measurement values taken within 10 minutes.
Notes:
Note: The cuff wrap guide result appears on the display with the measurement values. The cuff wrap guide lamp will not light.
: The arm cuff was wrapped snugly enough.
: The arm cuff was loose or not wrapped correctly.
Notes:
Note: You cannot partially delete the values stored in the memory.
All values for the user you select will be deleted.
Error Display | Cause | Solution |
Irregular heartbeats are detected. | Remove the arm cuff. Wait 2 - 3 minutes and then take another measurement. Repeat the steps in "Taking a Measurement" section. If this error continues to appear, contact your physician. | |
Movement during measurement. | Carefully read and repeat the steps in "Taking a Measurement" section. | |
The batteries are low. | You should replace the batteries with new ones ahead of time. | |
The batteries are exhausted. | You should replace the batteries with new ones at once. | |
Air plug disconnected. | Insert the plug securely. | |
Arm cuff is applied too loosely. | Apply the arm cuff tighter. | |
Air is leaking from the arm cuff. | Replace the cuff with a new one. | |
Movement during measurement and the arm cuff has not been inflated sufficiently. | Repeat measurement. Remain still and do not talk during measurement. | |
If "E2" appears repeatedly, inflate the cuff manually until it is 30 to 40 mmHg above your previous measurement result. | ||
The arm cuff was inflated above 299 mmHg when inflating the cuff manually. | Do not inflate the cuff above 299 mmHg. | |
Movement during measurement. | Repeat measurement. Remain still and do not talk during measurement. | |
Clothing is interfering with the arm cuff. | Remove any clothing interfering with the arm cuff. | |
Device error. | Contact your OMRON retail outlet or distributor. |
Problem | Cause | Solution |
The measurement result is extremely high (or low) | Arm cuff is applied too loosely. | Apply the arm cuff tighter. |
Movement or talking during measurement. | Remain still and do not talk during measurement. | |
Clothing is interfering with the arm cuff. | Remove any clothing interfering with the arm cuff. | |
Arm cuff pressure does not rise | The air connector is not securely connected into the air jack. | Make sure that the air tube is connected securely. |
Air is leaking from the arm cuff. | Replace the arm cuff with a new one. | |
The cuff wrapping guide lamp does not lightArm cuff deflates too soon | The arm cuff is loose. | Apply the cuff correctly so that it is firmly wrapped around the arm. |
Cannot measure or the results are too low or too high. | The arm cuff has not been inflated sufficiently. | Inflate the cuff so that it is 30 to 40 mmHg above your previous measurement result. |
Nothing happens | The batteries are empty. | Replace the batteries with new ones. |
The batteries have been inserted incorrectly. | Insert the batteries with the correct (+/-) polarity. | |
Other problems |
|
To protect your device from damage, please observe the following:
Keep the device in its storage case when not in use.
Do not store the device in the following situations:
(within the scope of EC Medical Device Directive 93/42/EEC)
To disconnect the AC adapter, unplug the AC adapter from the electrical outlet first and then remove the AC adapter plug from the monitor.
Product description | Automatic Blood Pressure Monitor |
Model | OMRON M3 (HEM-7131-E) |
Display | LCD Digital Display |
Measurement method | Oscillometric method |
Measurement range | Pressure: 0 to 299 mmHg Pulse: 40 to 180 beats/min. |
Accuracy | Pressure: ±3 mmHg Pulse: ±5% of display reading |
Inflation | Fuzzy-logic controlled by electric pump |
Deflation | Automatic pressure release valve |
Memory | 60 measurements with date and time for each user (1 and 2) |
Rating | DC6V 4W |
Power source | 4 "AA" batteries 1.5V or optional AC adapter (Adapter S-9515336-9, INPUT AC100-240V 50/60Hz 0.12A) (Adapter UK-9983666-5, INPUT AC100-240V 50/60Hz 15VA) |
Battery life | Approx. 1000 measurements (using new alkaline batteries) |
Applied part | = Type BF |
Protection against electric shock | Internally powered ME equipment (When using only the batteries) = Class II ME equipment (Optional AC adapter) |
Operating temperature/ humidity | +10 to +40°C / 30 to 85% RH |
Storage temperature/ humidity/ air pressure | -20 to +60°C / 10 to 95% RH / 700 to 1060 hPa |
IP classification | IP 20 |
Weight | Monitor: Approx. 280 g without batteries Arm cuff: Approx. 170 g |
Outer dimensions | Monitor: Approx. 107 (w) mm x 79 (h) mm x 141 (l) mm Arm cuff: Approx. 145 mm x 594 mm |
Cuff circumference | 22 to 42 cm |
Cuff/ Tube material | Nylon, polyester, polyvinyl chloride |
Package contents | Monitor, arm cuff, instruction manual, storage case, battery set, blood pressure pass |
Notes:
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC's and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation. Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to prevent unsafe product situations, the EN60601-1-2:2007 standard has been implemented. This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this EN60601-1-2:2007 standard for both immunity and emissions. Nevertheless, special precautions need to be observed:
Further documentation in accordance with EN60601-1-2:2007 is available at OMRON HEALTHCARE EUROPE at the address mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.
Correct Disposal of This Product (Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it should not be disposed of, with other household wastes at the end of its working life. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this product from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this product, or their local government office, for details of where and how they can return this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions of the purchase contract. This product should not be mixed with other commercial wastes for disposal.
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Advertisement
Thank you! Your question has been received!
Need Assistance?
Do you have a question about the M3 that isn't answered in the manual? Leave your question here.