Chapter 7 Specifi Cations; Specifi Cations - Omega DPF500 Series User Manual

Dpf500 series ratemeter/totalizer
Table of Contents

Advertisement

Chapter 7 Specifi cations
7.1 Specifi cations
Power:
BATTERY POWERED
Supplied with 1 or 2 C size Lithium battery pack.
EXTERNAL POWER INPUT
Voltage: 8.5 to 30 VDC
Current: Less than 5 mA
Supplied with 1 C size lithium battery
Protection: Reverse Polarity Protection on DC Power Input
LOOP POWERED
Voltage: 8.5 to 30 VDC
Supplied with 1 or 2 C size lithium battery(ies)
Protection: Reverse Polarity Protection on Current Loop
Loop Burden: 8.5V maximum
BATTERY LIFE EXPECTANCY:
Expected Years of Operation for DPF500 of various powering op-
tions at equipment duty cycles
MODEL
Idle
2hrs/day
-BAT*
-BAT*
10 yrs
10 yrs
10 yrs
10 yrs
-BAT-LIN
10 yrs
10 yrs
Standard unit
10 yrs
10 yrs
and -MA
(standby-operation)
Standard unit
Indefi nite operation when externally powered
and -MA
External or loop power
* -BAT option contains DPF5-BP battery pack
NOTE: Battery shelf life is rated at 10 years by manufacturer
Life expectancy based on rated battery capacity at 20°C
The above table is shown with pulse output inactive. Use
of pulse output shortens battery life.
Example: A pulse output of 0.06 sec. duration, once per
second, would derate the battery life by 20%.
DISPLAY:
Rate Display: (selectable decimal)
5 Digits (99999), 0.35" High, Display updates once per second
with battery power, 8X per second with DC or Loop power
Rate Descriptors:
/SEC, /MIN, /HR
Min. Input Frequency: 0.01 Hz to 10 Hz (selectable delay of 0.1
to 99.9 seconds)*
Selectable Rate Display Damping
Totalizer Display: (selectable decimal)
8 Digits (99999999), 0.2" High
Totalizer Descriptors:
GAL, LIT, FT3, M3, "blank"
Warning Displays: Low battery warning
PULSE OUTPUT:
The pulse output advances with the least signifi cant digit of the
totalizer or decimal multiples there of (see Pulse scale divider).
Type: Isolated photomos relay
Max. voltage (off state): 30 VDC
Current (on state): 100mA
Pulse Duration: Selectable 0.5, 0.25, 0.125, 0.0625 seconds
Pulse Scale divider (Pulscale): User selectable, ÷1, ÷10, ÷100
or OFF
NOTE: Select OFF for max. battery life.
ACCURACY:
0.01% Reading, ±1 count
Temperature Drift: 50 ppm/°C Worst Case
* A large delay setting and internal math operations may delay the update rate.
RUN TIME
8hrs/day
24hrs/day
10 yrs
10 yrs
9.1 yrs
9.1 yrs
10 yrs
8.4 yrs
10 yrs
10 yrs
ENVIRONMENTAL:
OPERATING TEMPERATURE
-4°F (-20°C) to + 158°F (70°C)
Extended Temp: -22°F (-30°C) to + 158°F (70°C)
HUMIDITY
0 - 90% Noncondensing
MOUNTING STYLES:
1- Panel Mount -
NEMA 4X Front
2- Wall Mount -
NEMA 4X Enclosure
(keypad mounted behind clear cover)
3- Explosion Proof -
Class I, Division I, Groups B, C & D
Class II, Division I, Groups E, F & G
INPUTS:
MAGNETIC PICKUP INPUT
Frequency Range: 0 to 3500 Hz
Trigger Sensitivity: 30 mV p-p (10 mV p-p on special order)
Over Voltage Protected: ± 30 VDC
OPTO-ISOLATED DC PULSE INPUT
High (logic 1): 4-30 VDC
Low (logic 0): Less Than 1 VDC
Minimum Current: .5 mA
Hysteresis: 0.4 VDC
Frequency Range: 0 to 5 kHz
Min. Pulse Width: 0.1 msec
CONTACT CLOSURE INPUT (contact closure to common)
Internal Pullup Resistor: 100 KΩ to +3.6 VDC
High (logic 1): Open or 4-30 VDC
Low (logic 0): Less Than .5 VDC
Internal Switch Debounce Filter: 0 to 40 Hz
NOTE:
Sustained contact closure will shorten
battery life.
RESET INPUT (contact closure to common)
Internal Pullup Resistor: 100 KΩ to +3.6 VDC
High (logic 1): Open or 4-30 VDC
Low (logic 0): Less Than .5 VDC
Minimum On : 25 msec
NOTE:
Sustained contact closure will shorten
battery life.
K-FACTOR
Range: 0.001 to 99999999
Decimal Point Locations: XXXX.XXXX to XXXXXXXX
20 POINT LINEARIZATION -LIN OPTION
This feature allows the user to enter 20 different frequencies
with 20 different corresponding K-Factors to linearize non linear
signals.
ANALOG OUTPUT -MA OPTION:
Type: 4-20 mA follows rate display, Two wire hookup
Accuracy: 0.025% Full Scale at 20° C
Temperature Drift: 50 ppm/°C Typical
Reverse Polarity Protected
Update Rate: 8 times/second
NOTE: The DPF500 uses the 4-20 mA loop power as its primary
power source when this option is used. The battery is still
required for standby battery operation.
DATA STORAGE:
Setup Information: Stored in fl ash memory
Totalizer: Stored in battery backed RAM but can be saved to
fl ash memory by operator for recall after battery change out.
12

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dpf501Dpf502Dpf503

Table of Contents