5.4 Disposable one-day pump set – Cleaning
1
Check whether there is milk in the
tubing. If so, replace the pump
set. If not, disassemble the pump
set into its individual parts (breast
shield, connector, membrane,
membrane cap, protective
membrane and tubing). You must
also remove the white valve from
the connector.
➙ Note that the membrane cap,
protective membrane and tubing
must not be cleaned.
5.5 Disposable one-day pump set
– Disinfection and sterilisation
Do not disinfect or sterilise the
disposable one-day pump set.
2
Rinse the breast shield, the connector
and the white valve with cold, clear water
(approx. 20 °C) to remove proteins. Clean
these parts with plenty of warm, soapy
water (approx. 30 °C) to wash off fat. Use a
commercially available washing-up liquid,
preferably without artificial fragrances and
colouring (5–15% anionic tenside, nonionic
tenside, pH neutral). Rinse the parts with
cold, clear water (approx. 20 °C).
5.6 Disposable one-day pump set –
Disposal
Dispose of the disposable one-day
pump set after a maximum of eight
pumping sessions within 24 hours.
21