RF emission
Class B
CISPR 11
Electromagnetic immunity
Guidance and manufacturer's declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS
The Blood pressure monitor (SBM 22) is intended for use in the electromagnetic environment specified below. The customer
of the user of the Blood pressure monitor (SBM 22) should assure that it is used in such and environment.
Immunity test
IEC 60601
test level
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Power frequency
3A/m
(50Hz) magnetic
field
IEC 61000-4-8
The Blood pressure monitor (SBM 22) is suitable for use in all establishments, inclu-
ding
domestic establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
Compliance
Electromagnetic environment – guidance
level
±6 kV contact
Floors should be wood, concrete or ceramic tile. If floor are
±8 kV air
covered with synthetic material, the relative humidity should be
at least 30%.
3A/m
Power frequency magnetic fields should be at levels characte-
ristic of a typical location in a typical commercial or hospital
environment.