Sanitas SBM 37 Instructions For Use Manual page 80

Hide thumbs Also See for SBM 37:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 17
Power frequency
(50Hz) magnetic field
IEC 61000-4-8
Guidance and manufacturer's declaration – electromagnetic immunity for EQUIPMENT and SYSTEMS that are not LIFE-
SUPPORTING
The Blood pressure monitor (SBM 37) is intended for use in the electromagnetic environment specified below. The customer
or the user of Blood pressure monitor (SBM 37) should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Conducted RF
3 V
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
3A/m
3A/m
Compliance
level
3 V
rms
rms
3 V/m
Power frequency magnetic fields should be at levels characte-
ristic of a typical location in a typical commercial or hospital
environment.
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should be used
no closer to any part of the Blood pressure monitor (SBM 37), inclu-
ding cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.
Recommended separation distance.
80 MHz to 800 MHz
800 MHz to 2.5 GHz
80