Nikon DS-2MBW Instruction Manual page 82

Camera control unit, camera head, cooled camera head
Hide thumbs Also See for DS-2MBW:
Table of Contents

Advertisement

.TOOL.
3 Measuring Two-point Distance or Entering Information—TOOL Menu
8)
Press
they have been saved.
NOTE: If you select "E" on the calibration screen, calibration values are converted
< When using the magnification of the objective and relay lenses >
1)
Select a number in which you want to save the calibration value from
Drag the mouse pointer over one of numbers
information for that number will be displayed in the Info area for your confirmation.
NOTE: If the microscope connected to your system is the Nikon 80i/90i microscope
2)
Press
3 )
Click the
keypad to enter the magnification of the objective.
4 )
Click the
keypad to enter the magnification of the relay lens.
5)
If you want to attach information to the calibration value you've created, click the Info
input box and enter text for the information you want to attach. This information will
prove helpful because you can confirm Info when you call up the calibration value at a
later time.
6)
Check the Set number to confirm that the calibration number is correct, then click
to save it.
7)
Press
NOTE:
*
*
VI Basic Operations
to return to the previous menu. Any calibration values will be erased unless
to inches or millimeters before saving. Values entered in centimeters or
millimeters are converted to inches, while values entered in micrometers or
nanometers are converted to millimeters.
with motorized or intelligent nosepiece, note that each calibration value call
number is linked to the hole number of the nosepiece. This means you
cannot change calibration values by selecting
is selected automatically from
number of the nosepiece you are currently using.
Calib.
to bring up the calibration screen shown below.
numeric value input box to display the numeric keypad. Use the
numeric value input box to display the numeric keypad. Use the
Save
is disabled if no value is entered.
to return to the previous menu. Any calibration values not saved are deleted.
If the microscope connected to your system is the Nikon 80i/90i microscope
with motorized or intelligent nosepiece and the calibration method you
selected is "using the magnification of the objective and relay lenses," you
do not need to enter the magnification of the objective. If the camera head
is connected to the rear port of the DIH, you do not need to enter the
magnification of the relay lens. If you calibrated with one objective, other
connected objectives will be also calibrated automatically with their
magnifications being converted.
If you selected the "using the magnification of the objective and relay
lenses" calibration method, keep in mind that the calibration value may
incorporate a certain amount of error attributable to optical errors of the
objective. With the Nikon 80i/90i microscope connected, errors may range
up to 10%.
s1
through
s7 , and the calibration
s1
through
s1
through
s7
based on the hole
- 72 -
s1
through
s7 .
s7 . The value
Save