10A AC/DC Current Measurements  
WARNING: Do not handle the test leads above the finger/hand guard barrier. 
CAUTION: Observe CAT III 600V with respect to Earth Ground. 
1.
Insert the black test lead into the COM terminal and the red test lead into the A terminal. 
2.
Turn the meter's function switch to the 
display indicating Amperes (Amps). 
3.
Press the M button to select AC or DC. 
4.
The meter defaults to the Auto Range mode. When in Auto mode, the display shows the 
Auto icon. Press the R button to manually range the meter. Press the R button to step 
through the ranges. Press and Hold the R button to return to the Auto Range mode. 
5.
Current measurements must be taken in series with the circuit under test. See 
accompanying diagram. 
6.
Touch the black test lead to the negative side of the circuit and the red test lead to the 
positive side of the circuit.  
7.
Read the current measurement in the display represented by numerical digits and bar 
graph. The display will indicate the proper decimal point and value. In DC, If the polarity 
is reversed, the display will show (‐) minus before the value.  
8.
For AC current, press the Hz/% button to view the frequency (Hz) or the duty cycle (%) of 
the measured current. 
 
% Duty Cycle Note: Duty Cycle measurements are 
AC only for this meter. Measurement signals must 
be 'zero crossing' and must reach ‐0.7V for the 
meter to properly display Duty Cycle %.
Fig 4‐3 AC/DC 10A CURRENT MEASUREMENTS 
 position. The A symbol will appear on the 
12
 
EX35x‐en‐GB_V1.2   12/15 