GE TuffSat User's Manual And Service Manual page 24

Table of Contents

Advertisement

TuffSat User's Guide and Service Manual
Message or
Cause
Condition
The sensor is not properly connected to
the oximeter.
Defective or incompatible sensor.
The connected sensor is not an OxyTip+
sensor.
Loose cable connection to the circuit
board.
Liquid in the oximeter sensor
connector.
Internal component malfunction.
The battery power is low—10 to 60
minutes remain for alkaline batteries;
Flashing on/off.
the time remaining for nonalkaline
batteries is not predictable.
Unexpected result
Debris (lint, etc.) is causing button to
when button is
stick.
pressed.
Damaged button cover.
Faulty circuit board.
3-4
Recommendation
Reconnect the sensor cable to the
sensor connector on the oximeter.
Replace the sensor.
Check the cable connection between the
sensor connector assembly and the
circuit board.
Disconnect the sensor and pour/shake
the liquid out of the sensor connector.
Replace the sensor connector assembly.
If the message reappears, replace the
circuit board.
Install new batteries.
Disassemble the oximeter and remove
debris from between the button cover
and the circuit board.
Replace the button cover.
Replace the circuit board.