Calibration - Health O Meter 3001KL-AM User Instructions

Antimicrobial digital platform scale
Hide thumbs Also See for 3001KL-AM:
Table of Contents

Advertisement

CALIBRATION

Calibration
This scale has been factory calibrated, and does not require calibration prior to use.
Calibration of your 3001KL-AM/3001KL-AMC scale can be performed using calibrated
kilogram (kg) or pound (lb) weight standards. The required calibration standard will be
the Unit of Measurement (UOM) for which the scale is configured.
Because the scale is shipped from the factory in "KG" mode, the requested UOM for
calibration will be the KG standard (e.g., 400kg). If the unit is to be calibrated with LB
weight standards, the scale must first be configured to the "LB" UOM. Refer to the Unit
of Measurement Configuration section for configuration details.
Warning: Always ensure that the scale is returned to the correct UOM after scale
calibration.
1. Ensure the display is powered off, then press and hold the
HOLD/RELEASE and ZERO button while pressing the ON/OFF
button.
2. Press the ENTER button.
3. Make sure there is nothing on the scale, then press the
ENTER button.
4. Use the ► and ▲ buttons to set the calibration weight to a
value other than the default weight. The calibration weight
should be as high as possible, with 270 KG/400 LB as a
minimum calibration weight.
5. Press the ENTER button.
6. Load the scale with the required weight.
7. Without touching or shaking the scale, allow the display to
stabilize, then press the ENTER button.
8. If the calibration is successful, the display will show SUCC
and power off. If the calibration is not successful, the display
will show Err1 and power off. Contact customer service for an
"Err1" display (+1 800-638-3722).
Operator Action
15
Display
4-5 digit 'idle scale count'
number.
"CAL".
"Clear" followed by "400.0 KG"
/ "0600.0 LB" if UOM is set to
LB.
Shows "400.0 KG" / "0600.0
LB" or entered calibration
weight.
Displays "Load" followed by
"400.0 KG" / "0600.0 LB".
Display increments showing
raw weight data.
"SUCC" or "Err1".
-------

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

3001kl-amc

Table of Contents