Electrical Specifications - Fluke 175 Calibration Information Manual

Table of Contents

Advertisement

Electrical Specifications

Function
[2] [3]
AC Volts
DC mV
DC Volts
Continuity
Ohms
Diode test
Capacitance
[5]
AC Amps
(True -rms)
(45 Hz to 1 kHz)
[5]
DC Amps
Hz
(AC- or DC- coupled,
[8] [9]
V or A
input )
-40 C to +400 C
[10]
Temperature
-40 F to +752 F
For dc functions, accuracy is the specified accuracy of the measurement function 12 counts for changes
longer than 350 ms in duration.
MIN MAX AVG
For ac functions, accuracy is the specified accuracy of the measurement function 40 counts for changes
longer than 900 ms in duration.
[1]
All ac voltage and ac current ranges are specified from 5 % of range to 100 % of range.
Crest factor of 3 at full scale up to 500 V, decreasing linearly to crest factor 1.5 at 1000 V.
[2]
[3]
For non-sinusoidal waveforms, add -(2 % reading + 2 % full scale) typical, for crest factors up to 3.
In the 9999 F range for measurements to 1000 F, the measurement accuracy is 1.2 % + 2 for all models.
[4]
[5]
Amps input burden voltage (typical): 400 mA input 2 mV/mA, 10 A input 37 mV/A.
[6]
400.0 mA accuracy specified up to 600 mA overload.
[7]
>10 A unspecified.
[8]
Frequency is specified from 2 Hz to 99.99 kHz in Volts and from 2 Hz to 30 kHz in Amps.
[9]
Below 2 Hz, the display shows zero Hz.
[10] In RF field of 3 V/m specified accuracy 5 C (9 F).
[11] Does not include error of the thermocouple probe.
[1]
Range
Resolution
600.0 mV
0.1 mV
6.000 V
0.001 V
60.00 V
0.01 V
600.0 V
0.1 V
1000 V
1 V
600.0 mV
0.1 mV
6.000 V
0.001 V
60.00 V
0.01 V
600.0 V
0.1 V
1000 V
1 V
600 
1 
600.0 
0.1 
6.000 k
0.001 k
60.00 k
0.01 k
600.0 k
0.1 k
6.000 M
0.001 M
50.00 M
0.01 M
2.400 V
0.001 V
1000 nF
1 nF
10.00 F
0.01 F
100.0 F
0.1 F
[4]
9999 F
1 F
60.00 mA
0.01 mA
[6]
400.0 mA
0.1 mA
6.000 A
0.001 A
[7]
10.00 A
0.01 A
60.00 mA
0.01 mA
[6]
400.0 mA
0.1 mA
6.000 A
0.001 A
[7]
10.00 A
0.01 A
99.99 Hz
0.01 Hz
999.9 Hz
0.1 Hz
9.999 kHz
0.001 kHz
99.99 kHz
0.01 kHz
0.1 C
0.1 F
Accuracy ( [ % of Reading ]   Counts ] )
175
1.0 % + 3
1.0 % + 3
(45 Hz to 500 Hz)
(45 Hz to 500 Hz)
2.0 % + 3
2.0 % + 3
(500 Hz to 1 kHz)
(500 Hz to 1 kHz)
0.15 % + 2
0.09 % + 2
0.15 % + 2
0.09 % + 2
0.15 % + 2
0.15 % + 2
Product beeps at <25 , beeper turns off at >250 ; detects
opens or shorts of 250 s or longer.
0.9 % + 2
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
1.5 % + 3
1 % + 2
1.2 % + 2
1.2 % + 2
1.2 % + 2
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
10 % typical
1.5 % + 3
1.5 % + 3
1.0 % + 3
1.0 % + 3
0.1 % + 1
0.1 % + 1
NA
True-rms Multimeters
Electrical Specifications
177
179
1.0 % + 3
(45 Hz to 500 Hz)
2.0 % + 3
(500 Hz to 1 kHz)
0.09 % + 2
0.09 % + 2
0.15 % + 2
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
1.5 % + 3
1.0 % + 3
0.1 % + 1
[11]
1 % + 10
NA
[10]
1 % + 18
5