Fluke 175 Calibration Information Manual page 14

Table of Contents

Advertisement

175, 177, 179
Calibration Information
Test (Switch Position)
[2]
Ohms
[1]
Capacitance
Continuity
[1]
Diode
AC Milliamps
AC Amps
m
[1]
DC Milliamp
[1]
DC Amps
[1] [3] [4]
Temperature
(179 only)
[1] Press  to access this function.
[2] Does not include test lead resistance.
[3] Be sure to use correct cable.
[4] To ensure accuracy, the Product and thermocouple adapter must be at the same temperature. Allow 20 minutes before
performing the temperature procedures.
14
Table 4. Performance Tests 177/179 (continued)
5500 Output
500 
5 k
50 k
500 k
5 M
19 M
Apply 0.90 F
Apply 9 F
Apply 90 F
Apply 900 F
25 
250 
2.0 V
3 mA 45 Hz
50 mA 1 kHz
400 mA 1 kHz
4.0 A 45 Hz
9.0 A 1 kHz
3 mA
50 mA
-400 mA
4.0 A
-9.0 A
0 C
-40 C
400 C
Product Response
Lower Limit
495.3 
4.954 k
49.54 k
495.4 k
4.954 M
18.68 M
887 nF
8.87 F
88.7 F
887 F
Beeper On
Beeper Off
1.978 V dc
2.92 mA ac
49.22 mA ac
393.7 mA ac
3.937 A ac
8.83 A ac
2.94 mA dc
49.47 mA dc
-404.3 mA dc
3.957 A dc
-9.12 A dc
-1.0 C
-41.4 C
395.0 C
Upper Limit
504.7 
5.046 k
50.46 k
504.6 k
5.046 M
19.32 M
913 nF
9.13 F
91.3 F
913 F
2.022 V dc
3.08 mA ac
50.78 mA ac
406.3 mA ac
4.063 A ac
9.17 A ac
3.06 mA dc
50.53 mA dc
-395.7 mA dc
4.043 A dc
-8.88 A dc
1.0 C
-38.6 C
405.0 C