175, 177, 179
Calibration Information
The Product is damaged and requires service if:
and messages continue to appear after a proper recalibration.
and messages are alternating on the display.
message appears on the display.
If the Product has a serial number of 84060000 or below, or 83911241 to 84060000, use the following steps.
Function (Switch Position)
(AC Volts)
(DC Volts)
DC Millivolts
(Temperature - 179 only)
(Ohms)
(Continuity)
mA
(Milliamps)
A
(Amps)
[1] To ensure accuracy, the Product and Thermocouple Adapter must be at the same temperature. Allow 20 minutes before
performing the temperature procedures.
16
Table 5. Calibration Steps
Calibration Step
C-01
C-02
C-03
C-04
C-05
C-06
C-07
C-08
C-09
[1]
C-09 (175/177)
C-10 (179)
C-10 (175/177)
C-11 (179)
C-11 (175/177)
C-12 (179)
C-12 (175/177)
C-13 (179)
aik14f2.eps
Input Value
6.000 V, 900.0 Hz
600.0 mV, 100 Hz
6.000 V, 100 Hz
6.000 V dc
60.00 V dc
600.0 V dc
1000 V dc
600.0 mV dc
0.0 C
6.000 k
600
600.0 mA dc
6.000 A dc