Fluke 175 Calibration Information Manual page 13

Table of Contents

Advertisement

Test (Switch Position)
AC Volts
Hz
[1]
AC Volts Frequency
DC Volts
 Hz
[1]
DC Volts Frequency
DC Millivolts
Table 4. Performance Tests 177/179
5500 Output
300 mV 45 Hz
500 mV 500 Hz
500 mV 1 kHz
5 V 45 Hz
5 V 500 Hz
5 V 1 kHz
50 V 45 Hz
50 V 500 Hz
50 V 1 kHz
300 V 45 Hz
500 V 500 Hz
500 V 1 kHz
1000 V 45 Hz
1000 V 500 Hz
1000 V 1 kHz
1 V 45 Hz
1 V 900 Hz
1 V 9 kHz
5 V 50 kHz
5 V
50 V
300 V
1000 V
-1000 V
3 V 45 kHz
30 V 50 kHz
30 mV 0 Hz
-300 mV
600 mV
True-rms Multimeters
Product Response
Lower Limit
296.7 mV ac
494.7 mV ac
489.7 mV ac
4.947 V ac
4.947 V ac
4.897 V ac
49.47 V ac
49.47 V ac
48.97 V ac
296.7 V ac
494.7 V ac
489.7 V ac
987 V ac
987 V ac
977 V ac
44.94 Hz
899 Hz
8.99 kHz
49.94 kHz
4.993 V dc
49.93 V dc
299.5 V dc
997 V dc
-1003 V dc
44.94 Hz
49.94 kHz
29.8 mV dc
-300.5 mV dc
599.3 mV dc
Performance Tests
Upper Limit
303.3 mV ac
505.3 mV ac
510.3 mV ac
5.103 V ac
5.053 V ac
5.103 V ac
50.53 V ac
50.53 V ac
51.03 V ac
303.3 V ac
505.3 V ac
510.3 V ac
1013 V ac
1013 V ac
1023 V ac
45.06 Hz
901 Hz
9.01 kHz
50.06 kHz
5.007 V dc
50.07 V dc
300.5 V dc
1003 V dc
-997 V dc
45.06 Hz
50.06 kHz
30.2 mV dc
-299.5 mV dc
600.7 mV dc
13