10 MHz Module (4395B-10)
This module provides four 10 MHz outputs. The module also monitors each output for levels
below approximately 10 dBm, after which an alarm is raised.
Characteristic
Number of outputs
Output connectors
Impedance
Amplitude
Signal
Harmonics
Phase Noise
5 MHz Module (4395A-5)
This module provides four 5 MHz outputs. The module also monitors each output for levels below
approximately 10 dBm, after which an alarm is raised.
Characteristic
Number of outputs
Output connectors
Impedance
Amplitude
Signal
Harmonic
Phase Noise (dBc/Hz)
1 MHz
-150 dBc/Hz
Four
BNC(F)
50
+13dbm +/- 2dBm
Sine Wave
< -40dBc
ATS 6511C
0.1Hz
-80
dBc/Hz
1 Hz
-110 dBc/Hz
10 Hz
-132 dBc/Hz
100 Hz
-136 dBc/Hz
1 kHz
-141 dBc/Hz
10 kHz
-141 dBc/Hz
100 kHz -155 dBc/Hz
1 MHz
-155 dBc/Hz
Four
BNC(F)
50
+13 dbm ± 2 dBm
Sine Wave
< -40dBc
ATS 6511A/B
1 Hz
-106 dBc/Hz
10 Hz
-131 dBc/Hz
100 Hz
-138 dBc/Hz
1 kHz
-144 dBc/Hz
10 kHz
-146 dBc/Hz
100 kHz -146 dBc/Hz
1 MHz
-146 dBc/Hz
62
DOC 6511_Release K
ATS-6511A/B/C Users Guide
-150 dBc/Hz
Description
Description
ATS 6511C
-116 dBc/Hz
-138 dBc/Hz
-140 dBc/Hz
-144 dBc/Hz
-146 dBc/Hz
-146 dBc/Hz
-146 dBc/Hz