SECTION 6—OPERATING INSTRUCTIONS
Operating Alarm – If unit detects abnormal operating
conditions in the unit, the unit will alarm with a rapid audible
beep, the red alarm indicator light will illuminate
continuously, the unit will not operate and the flow setting 3
& 4 blue indicator lights will illuminate. Turn Off and then
turn On the unit. If the alarm continues, you should change to
another source of oxygen and contact your equipment
provider.
Compressor Alarm – If unit detects abnormal compressor
conditions, the unit will alarm with a rapid audible beep, the
red alarm indicator light will illuminate continuously, the unit
will not operate and the flow setting 4 & 5 blue indicator
lights will illuminate. Turn Off and then turn On the unit. If
the alarm continues, you should change to another source of
oxygen and contact your equipment provider.
System Alarm – If unit detects abnormal system conditions,
the unit will alarm with a rapid audible beep, the red alarm
indicator light will illuminate continuously, the unit will not
operate and the flow setting 3, 4 & 5 blue indicator lights will
illuminate. Turn Off and then turn On the unit. If the alarm
continues, you should change to another source of oxygen
and contact your equipment provider.
Audible Alarms - To temporarily silence the audible alarms,
quickly press the On/Off button. Except for the system Too
Hot/Too Cold alarms, this feature will reset itself, every 30
seconds, each time a new alarm is triggered, or if the unit is
turned Off. When performed during the system Too Hot/Too
Cold alarms, the audible alarm will be silenced until a new
alarm is triggered, or the unit is turned Off.