Location; Recommendations For Optimal Performance - Invacare XPO2 Operator's Manual

Portable concentrator
Hide thumbs Also See for XPO2:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
SECTION 6—OPERATING INSTRUCTIONS

Location

NEVER block the air openings of the product or place it on a
soft surface, such as a bed or couch, where the air opening
may be blocked. Keep the openings free from lint, hair and the
like.
Keep unit at least 3 in (7.6 cm) away from walls, draperies,
furniture, and the like.
Locate and position the XPO
the air intake and the air exhausts are not obstructed.
Recommendations for Optimal
Performance
Operating Temperature:
Relative Humidity:
Transport/Storage
Temperature:
Electrical:
Altitude:
Tubing and Cannula:
Environment:
Time of Operation:
Invacare® XPO
2
WARNING
in a well ventilated space so that
2
41°F to 95°F (5°C to 35°C)
15% to 60%
-2°F to 140°F (-20°C to 60°C)
Allow unit to warm, or cool, to operating
temperature range before using.
No extension cords.
Up to 10,000 ft (3046 m) above sea level.
4 ft (1.2 m) cannula with a maximum 25 ft
(7.6 m) of crush resistant tubing (DO NOT
pinch).
Smoke, pollutant and fume free. No
confined spaces (example: no closets).
Up to 24 hours per day when connected to
AC or DC power adapter.
18
Part No 1160887

Advertisement

Table of Contents
loading

Table of Contents