Bayer HealthCare CONTOUR NEXT LINK User Manual page 9

Wireless blood glucose monitoring system
Hide thumbs Also See for CONTOUR NEXT LINK:
Table of Contents

Advertisement

High and Low Alerts
The High and Low Alert feature tells you when your test result
is above or below the settings you choose. Results above High
are marked High Blood Sugar. Results below Low are marked
Low Blood Sugar. Alerts appear as large orange numbers.
NOTE: Alerts can only be set outside your selected
Target Range(s). For additional information, you may
contact Customer Service. Contact information for
Customer Service is on the back cover.
WARNING
Talk to your health care professional before setting any
High or Low Alerts on your meter.
Your meter comes preset with a high alert level of 13.9 mmol/L
and a low alert level of 3.9 mmol/L. You can Accept or Change
these levels.
HIGH AND LOW ALERTS
mmol/L
High: 13.9
Accept
mmol/L
Low: 3.9
Change
NOTE: Your Target ranges must be within your High and
Low Alert levels. You set your Target ranges next.
HIGH AND LOW ALERTS
mmol/L
High:
13.9
OK
mmol/L
Low: 3.9
®
10
19. Select Accept to keep the
preset alert levels. If you
wish to change these
alert levels, press Change.
20. If you selected Change,
press the L or M
button to select your
High Alert level. This
can be set between
7.0 - 20.0 mmol/L.
Press OK.
HIGH AND LOW ALERTS
mmol/L
High: 14.0
OK
mmol/L
Low:
3.9
NOTE: Press and hold the L or M button to scroll
through the numbers faster.
Alerts are set.
mmol/L
High: 14.0
Accept
mmol/L
Change
Low: 3.8
Target Ranges
Next you are asked to accept your blood sugar Targets. Targets
are your personal ranges for your blood sugar results. With
AutoLog OFF, you have a single Target range. With AutoLog
ON, you have Target ranges for Fasting, Before Meal, and
After Meal.
NOTE: Targets can only be set inside your selected Alert
level(s). For additional information, you may contact
Customer Service. Contact information for Customer
Service is on the back cover.
WARNING
Talk to your health care professional before setting any
Target ranges on your meter.
TARGET
Accept
3.9 - 10.0
Change
21. Press the L or M
button to select your
Low Alert level. This
can be set between
3.0 - 5.5 mmol/L.
Press OK.
22. The next screen confirms
your choices. Press
Accept to confirm your
Alert settings.
23. If AutoLog is OFF, Accept
or Change the preset
blood sugar Target
range. The preset range
is 3.9 - 10.0 mmol/L.
11

Advertisement

Table of Contents
loading

Table of Contents