  • Page 1 Wireless Wheelchair Scale Model D300 Series Version 1.0 Operation Manual PN 161817...
  • Page 3: Table Of Contents

    Course descriptions and dates can be viewed at www.ricelake.com/training or obtained by calling 715-234-9171 and asking for the training department. © Rice Lake Weighing Systems. All rights reserved. Printed in the United States of America. Specifications subject to change without notice.
  • Page 4 Rice Lake continually offers web-based video training on a growing selection of product-related topics at no cost. Visit www.ricelake.com/webinars. D300 Series Wireless Wheelchair Scale Operation Manual...
  • Page 5: About This Manual

    The D300 Series wireless wheelchair scale provides home healthcare, skilled nursing facilities, and hospital communities with a lightweight, portable and wireless alternative to the critical task of weighing wheelchair bound patients. The wireless wheelchair scale is easy to set up and operate. The D300 Series comes in two models: •...
  • Page 6: Fcc/Ce Notice

    The device must not be cleaned with lubricant or alcohol-based products. Plate electrode and coating can be peeled off. Use a dry cloth or wet tissue. CAUTION In case of mechanical failure, please contact Rice Lake Weighing Systems. This equipment has been tested and found to comply with the limits for medical devices to the IEC Important 60601-1-2:1994.
  • Page 7: System Setup

    System Setup Unpacking the Scale Visually inspect the D300 Series scale to ensure all components are included and undamaged. Standard hardware includes: Item Description Quantity Scale Pads Handheld Controller and Rubber Glove for Controller 1 each Roll on Ramps Lanyard Wheel Stop Table 2-1.
  • Page 8: Repacklng

    Repacklng If the D300 Series scale must be returned for repair, it must be properly packed with sufficient packing materials. Whenever possible, use the original carton when shipping the scale back. Important Damage caused by improper packaging is not covered by the warranty.
  • Page 9: Using The D300 Series

    Using the D300 Series The D300 Series wireless wheelchair scale has two different modes available when using the scale. • Weigh Patient - gives a displayed weight of the patient and wheelchair. • Weigh Patient & Chair - gives the weight of the patient without the weight of the wheelchair.
  • Page 10 8. Place the ramps onto the scale pad (shown in Figure 3-5). Wheel Stop Ramp Figure 3-5. Wheel Stop and Ramp Attached to Scale Pad D300 Series Wireless Wheelchair Scale Operation Manual...
  • Page 11: Weigh Patient And Chair

    There are two variations that can be done while in this mode: • Chair • Patient + Chair Main Menu Weigh Patient Weigh Patient & Chair Zero Out Pads Figure 3-7. Weighing Patient & Chair D300 Series Wireless Wheelchair Scale Operation Manual — Using the D300 Series...
  • Page 12 Note Once the hand held controller is turned off, the wheelchair weight is gone. 6. Attach ramps back onto the scale pads. 7. Release wheelchair brakes and remove wheelchair off of scale pads. D300 Series Wireless Wheelchair Scale Operation Manual...
  • Page 13 Continue Figure 3-12. Zeroing out Pads 6. Press to initiate the zero out function. (When the controller is in process, the screen will go dim for continue D300 Series Wireless Wheelchair Scale Operation Manual — Using the D300 Series...
  • Page 14 Press to return back to the main menu. 14. Attach ramps back onto scale pads. 15. Remove wheelchair and patient off of scale pads. D300 Series Wireless Wheelchair Scale Operation Manual...
  • Page 15: Options Menu

    Options Menu There are several options available on the D300 Series wireless wheelchair scale. To access those options, press the upper-right display button on the hand held controller. Setup/Options Button Options Help Beep Preferences Date & Time Deep Settings Figure 3-14. Setup/Options Button Location 3.3.1...
  • Page 16: Troubleshooting

    If all directions have been followed and are still not getting the desired results, call (800) 472-6703 and ask for technical support. • Yellow and red indication lights mean inconsistent weight readings. Ensure scale pads are placed on level flooring, such as tile, cement or hardwood. D300 Series Wireless Wheelchair Scale Operation Manual...
  • Page 17: D300 Series Limited Warranty

    D300 Series Limited Warranty Rice Lake Weighing Systems (RLWS) warrants that all RLWS equipment and systems properly installed by a Distributor or Original Equipment Manufacturer (OEM) will operate per written specifications as confirmed by the Distributor/OEM and accepted by RLWS. All systems and components are warranted against defects in materials and workmanship for one year.
  • Page 18 D300 Series Wireless Wheelchair Scale Operation Manual...
  • Page 20 230 W. Coleman St. Rice Lake, WI 54868 USA U.S. 800-472-6703 Canada/Mexico 800-321-6703 International 715-234-9171 Europe +31 (0) 88 2349171 www.ricelake.com www.ricelake.mx www.ricelake.eu www.ricelake.co.in m.ricelake.com © Rice Lake Weighing Systems 04/2015 PN 161817...