Recommended Equipment
5500A Multi-Product Calibrator
(or equivalent)
5500A Multi-product Calibrator
(or equivalent)
8
Table 2. Required Equipment
Measurement Function
DC Volts
DC Current
AC Volts
AC Current
Resistance
Capacitance
Frequency
Accuracy
10 mV to 600 V
±0.125 %
600 μA to 10 A
±0.25 %
6 mV to 600 V
±0.25 % @ 45 Hz to 1 kHz
600 µA to 10 A
±0.375 % @ 45 Hz to 1 kHz
0 Ω to 5 MΩ
±0.225 %
10 Ω to 30 MΩ
±0.375 %
9 μF to 900 μF
±0.475 %
2 V @ 50 kHz
±0.025 %