Fluke CNX 3000 Calibration Manual page 21

Hide thumbs Also See for CNX 3000:
Table of Contents

Advertisement

Test
(Switch
Input
Position)
.01 V
-5 V
50 V
DC Volts
10 V
-500 V
1000 V
3 mV
500 mV
DC Millivolts
-500 mV
30 mV 60 Hz

500 mV 60 Hz
AC Millivolts
500 mV 1 kHz
6 Ω
500 Ω
.6 kΩ
5 kΩΩ
Ohms
50 kΩ
500 kΩ
5 MΩ
50 MΩ
10 nF
900 nF
9 μF
Capacitance
90 μF
900 μF
24 Ω
251 Ω
Continuity
1.95 V
Diode
50 mA 60 Hz
mA AC
350 mA 1 kHz
50 mA
DC Amps
350 mA
Table 3. Performance Tests (cont.)
Lower Limit
0.008
-5.006
49.94
9.7
-500.6
997
2.8
499.4
-500.6
29.4
494.7
489.7
5.8
497.3
.596
4.974
49.74
497.4
4.974
49.22
8
887
8.87
88.7
887
Beeper On
Beeper Off
1.928
49.22
344.4
49.72
347.9
Upper Limit
0.012
-4.994
50.06
10.3
-499.4
1003
3.2
500.6
-499.4
30.6
505.3
510.3
6.2
502.7
.604
5.026
50.26
502.6
5.026
50.78
12
913
9.13
91.3
913
1.972
50.78
355.6
50.28
352.1
Wireless Multimeter
Performance Tests
Units
V dc
mV dc
V dc
mV ac
Ω
nF
μF
V dc
mA ac
mA dc
13

Advertisement

Table of Contents
loading