Electromagnetic compatibility

The AirCurve 10 complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-
2:2007, for residential, commercial and light industry environments. It is recommended that mobile communication devices
are kept at least 1 m away from the device.
Information regarding the electromagnetic emissions and immunity of this ResMed device can be found on
www.resmed.com, on the Products page under Service and Support.
EN 60601-1:2006 classification
Class II (double insulation), Type BF, Ingress protection IP22.

Sensors

Pressure sensor:
Flow sensor:

Maximum single fault steady pressure

Device will shut down in the presence of a single fault if the steady state pressure exceeds:
30 cm H
O (30 hPa) for more than 6 sec or 40 cm H
2
Sound
Pressure level measured according to EN ISO 17510-1:2009 (CPAP mode):
SlimLine:
Standard:
Power level measured according to EN ISO 17510-1:2009 (CPAP mode):
SlimLine:
Standard:
Declared dual-number noise emission values in accordance with ISO 4871:1996.
Physical - device and humidifier
Dimensions (H x W x D):
Air outlet (complies with ISO 5356-1:2004):
Weight (device and cleanable humidifier):
Housing construction:
Water capacity:
Cleanable humidifier - material:

Temperature

Maximum heater plate:
Cut-out:
Maximum gas temperature:
Air filter
Standard:
Hypoallergenic:
Aircraft use
ResMed confirms that device meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21,
category M) for all phases of air travel.
16
Internally located at device outlet, analogue gauge pressure
type, -5 to +45 cm H
Internally located at device inlet, digital mass flow type, -70
to +180 L/min
O (40 hPa) for more than 1 sec.
2
26.6 dBA with uncertainty of 2 dBA
26.6 dBA with uncertainty of 2 dBA
34.6 dBA with uncertainty of 2 dBA
34.6 dBA with uncertainty of 2 dBA
116 mm x 255 mm x 150 mm
22 mm
1248 g
Flame retardant engineering thermoplastic
To maximum fill line 380 mL
Injection moulded plastic, stainless steel and silicone seal
68°C
74°C
 41°C
Material: Polyester non woven fibre
Average arrestance: >75% for ~7 micron dust
Material: Acrylic and polypropylene fibres in a polypropylene
carrier
Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron
dust
O (-5 to +45 hPa)
2

Hide quick links: