Maintenance Instructions - Leica DM300 User Manual

Table of Contents

Advertisement

Maintenance instructions

General Notes
Protect the microscope from moisture,
Ϙ
fumes and acids and from alkaline, caustic
and corrosive materials. Do not store
chemicals in the vicinity.
Protect the microscope from oil and grease.
Ϙ
Never grease or oil mechanical parts or
sliding surfaces.
Follow the instructions of the disinfectant
Ϙ
manufacturer.
We recommend concluding a service
Ϙ
contract with Leica Service.
Leica DM300
Cleaning coated Parts and plastic Parts
Dust and dirt particles should be removed
Ϙ
with a soft brush or lint-free cotton cloth.
Remove coarse debris with a moistened
Ϙ
disposable cloth.
Acetone, xylene or nitro-containing thin-
Ϙ
ners must NOT be used.
Never use chemicals to clean colored
Ϙ
surfaces or accessories with rubberized
parts. This could damage the surfaces, and
specimens could be contaminated by abra-
ded particles.