Connecting The Electrical Power - Getinge 600 Series Installation Manual

Steam sterilizers
Hide thumbs Also See for 600 Series:
Table of Contents

Advertisement

Optional Equipment
CONNECTING THE ELECTRICAL
POWER
115 Vac ±10%
Single-Phase Supply
10A
UNINTERRUPTIBLE
POWER SUPPLY
(115 Vac Model)
230 Vac ±10%
Single-Phase Supply
10A
UNINTERRUPTIBLE
POWER SUPPLY
(230 Vac Model)
5–20
To connect the UPS:
1. Prepare the UPS cables as described in the UPS Installation
Instructions (Pub. No. 61301605360).
NOTE
The UPS cables are marked with their destination (e.g., 120V Supply, 1A
Power Box). The cable that connects to the sterilizer is fitted with a liquid-
tight connector.
2. Connect the UPS to the sterilizer:
a. If the mains voltage is 115 Vac, connect the UPS directly to Power
Box 1A (see Figure 5–16).
FIGURE 5–16. UPS CONNECTION (115 VAC)
b. If the mains voltage is 230 Vac, connect the UPS to Integral
Transformer Box 9A (see Figure 5–17).
FIGURE 5–17. UPS CONNECTION (230 VAC)
230 Vac
L
N
9A
1
2 3
1TB
INTEGRAL
1
2 3
2TB
TRANSFORMER
BOX
1A
POWER BOX
1A
POWER BOX
115 Vac
INS 61301607027 Rev A US
1TB
1
2
3
L
N
A08205-A1
1TB
1
2
3
L
N
A08205-A2