CONNECTING THE
TRANSFORMER
INS 61301607027 Rev A US
To connect the transformer:
1. After the transformer is installed, make the mains power supply
connections (L1 & L2) at the transformer terminal block as shown in
Figure 5–13 on page 5–16.
2. Wire the primary taps of the transformer for the correct voltage (See
chart located on the inside cover of the transformer).
3. Make the electrical connections from the sterilizer power box (1A) to
the transformer (8A) as shown in Figure 5–13 on page 5–16.