UNIVERSAL TRANSFORMER

Customer Supply
Universal
Transformer
8A
Output to Power Box
(115 Vac)
380/400/415/440/460/480 Vac
8A
UNIVERSAL
TRANSFORMER
5–16
When the customer power supply is 380 Vac or above, a transformer is
required to step down the voltage to the 115 Vac required by the sterilizer
controls. The customer is to provide:
• mounting of the transformer box.
• a connection from the transformer to the sterilizer power box.
• a connection from the customer supply (mains) to the transformer that
meets local code requirements.
• the required fused disconnects.
CAUTION
If a UPS will be installed, see UPS Installation Instructions (Pub. No.
61301605360) for electrical connections.
FIGURE 5–13.
Single Phase Supply
L1 L2
1
2
3
4
1TB
POWER SUPPLY CONNECTIONS
for Customer Supply Voltage
1A
POWER BOX
115 Vac
Wire Transformer Taps
(see Diagram Inside Box)
1TB
1
2
3
L
N