SAFETY FEATURES
DOOR LOCKOUT
DOOR OBSTRUCTION SWITCH
Top of Door
INS 61301607027 Rev A US
WARNING
SHOCK HAZARD: To disconnect all power, turn OFF the sterilizer at the
mains circuit breaker. The On/Off Controls do not remove primary power
from the sterilizer.
NOTE
The maximum sound pressure level of the equipment that the operator is
exposed to does not exceed 85 dBA
The door can be locked in the open position for maintenance when inside
the chamber. Turn the key to DISABLE and remove it. To prevent someone
from closing the door during maintenance, keep the keys with you.
FIGURE 2–1. LOCKING THE DOOR IN THE OPEN POSITION
WS-0420
The sterilizer uses a light beam switch at the top of the door to sense when
an object is in the path of the movement of the door. When the light beam
is broken as the door is closing, the door stops and reverses for two
seconds.
FIGURE 2–2. DOOR OBSTRUCTION SWITCH
Door Obstruction Switch
600 Series Steam Sterilizers
WS-0421
Light Beam