INSTALLING THE INTERIOR EQUIPMENT

INSTALLING THE RACK AND
SHELVES
1
Shelf Guide
INS 61301607027 Rev A US
The interior equipment package includes a left-hand shelf support, right-
hand shelf support, and two shelves. One or two additional shelves are
available as an option.
To install the rack and shelves:
1. Align the holes in the uprights of the shelf support with the mounting
studs on the side wall of the chamber.
2. Secure the shelf support to the mounting studs with the hardware
provided.
3. Repeat the procedure for the other shelf support.
NOTE
On 26-inch units, check that the shelf edge facing out has shelf guides.
FIGURE 4–28. INSTALLING A SHELF
Shelf Guide
3
Stop Tab
4. Set the shelf on top of the shelf support brackets.
5. Slide the shelf back until the back edge drops down at the end of the
support bracket.
6. Pull the shelf forward until the shelf guide contacts the stop tab on the
shelf support to be sure it is properly installed. Then slide the shelf to
the back of the chamber.
600 Series Steam Sterilizers
Outward Facing Shelf Edge
2
WS-0218-633HC
Shelf Guide
4–31