LEVELING THE STERILIZER
DOOR
0.25 in (6.3 mm)
(bottom of plate to finished floor)
INS 61301607027 Rev A US
Once the sterilizer is leveled, check the alignment of the door mechanism
to ensure that the door is level and the position of the mounting plate has
not changed.
To level the sterilizer door:
1. Remove the rubber bumpers from the door motor mounting plate. This
will allow a 9 inch torpedo level to be placed on the mounting plate.
2. Adjust the mounting plate leveling bolts until the bottom of the plate is
0.25 in. (6.3 mm) above the finished floor.
FIGURE 4–13. MOUNTING PLATE LEVELING BOLTS
Adjust to level plate
from side to side
3. Position a level as shown in Figure 4–13 and check that the mounting
plate is level from side to side.
600 Series Steam Sterilizers
Torpedo Level
Remove bumper to
make space for level
A07027-BE
4–13