Getinge 600 Series Installation Manual page 73

Steam sterilizers
Hide thumbs Also See for 600 Series:
Table of Contents

Advertisement

Customer Supply
115 or 230 Vac
Single-Phase
or
(G)
380/400/415 Vac
1TB
440/460/480 Vac
L1
1
Single-Phase
2
L2
3
(BK)
4
(W)
8A
Universal
Transformer
INS 61301607027 Rev A US
FIGURE 5–12. BOOSTER PUMP ELECTRICAL CONNECTIONS
From 2A6PO
(Water to Ejector)
Customer Supply
Current
Adjustment
Customer
Supply Wires
1K4
2A6PO4 (BK)
4 1K96
3
2
2A6PO1 (R)
1KA1
1
1J
Customer Supply
(115 or 230 Vac Single-Phase)
1J
2J
to
Water to Ejector
Solenoid Valve
6SV
11A
Pump I/F Box
1K
11AGND(G)
5
5
6
6
1B(W)
11A1K6(W)
3
3
4
4
1K5
2
1
2
1
11A1K2(BK)
1B(BK)
95
A1
A2
96
6SV(W)
4
1K96
3
2
6SV(BK)
1KA1
1
2J
600 Series Steam Sterilizers
Pump Motor
Contactor 1K
Overload
Relay
to
Booster Pump
1B
Front View of Box
w/o Cover
Water Booster
Pump
1B
Control Box
To
Water to Ejector
Solenoid Valve
6SV
2A
6JO
1
2
3
4
2A6PO
A08205-G
5–15

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents