POSITIONING THE STERILIZER

INS 61301607027 Rev A US
To position the sterilizer:
1. Roll the sterilizer into its final location.
FIGURE 4–5. LEVELING FEET ADJUSTMENT
2. Refer to Figure 4–5. From the underside of the sterilizer frame, rotate
the leveling feet in the frame until the bottom of the foot block is
approximately 2.25 inches (57 mm) above the floor (all four corners).
NOTE
This clearance is required for the installation of the kick panel(s) and
optional cabinet panels.
3. IMPORTANT: If seismic anchorage is required by local codes:
a. Position the seismic anchoring brackets on the leveling feet (see
Figure 4–25 on page 4–28).
b. Using the brackets as a template, mark the holes.
c. Raise the leveling feet and roll the sterilizer out of the way.
d. Drill the holes and install the anchoring devices into the floor. See
the seismic anchorage drawing on page 6–10.
e. Roll the sterilizer back into position.
f. Lower the leveling feet.
600 Series Steam Sterilizers
Foot Block
2.25 in. (57 mm)
from finished floor
A07027-BY
4–5