Getinge 600 Series Installation Manual page 32

Steam sterilizers
Hide thumbs Also See for 600 Series:
Table of Contents

Advertisement

Installation Instructions
4–6
NOTE
If a loading car and transfer carriage will be used to load goods in the
sterilizer, the tracks inside the chamber must match the height of the top
plate of the transfer carriage (see Figure 4–6 on page 4–6).
The standard height of the top plate of transfer carriages is 31.50 in. (800
mm) above the floor. Because a transfer carriage may be used with several
sterilizers, adjust the sterilizer height to match the transfer carriage.
FIGURE 4–6. TRANSFER CARRIAGE HEIGHT
Loading
Car
Transfer
Carriage
31.50 in. (800 mm)
Finished Floor
A07026-BR