Setting Up Time Blocks: Important Information
• Time blocks cover a 24-hour time period (from midnight to midnight).
• You can change the time period for any of the preset time blocks.
• When setting up bolus advice, you must complete and save the settings in at least one time block.
• Each time block must be at least 30 minutes long and can only be set in 30-minute increments.
• When you set the end time for a time block, the meter automatically sets this end time as the start time for
the next time block.
• You can set a different target blood glucose range, carb ratio, and insulin sensitivity for each time block.
• The target range for each time block must be within the hypo and hyper warning limits.
• When a time block is edited for the first time, the settings (target range, carb ratio, and insulin sensitivity)
are applied to all of the other time blocks.
Health Events
Health events allow you to select settings to take into account various activities or events that increase or
decrease your insulin need. Health events can be selected to indicate how you are feeling or what you are
doing that might affect your diabetes. The meter allows you the option of setting a percentage for each health
event except for Fasting. Fasting is not adjustable and does not scale bolus advice calculations. When a health
event is selected, the set percentage is used to increase or decrease the bolus insulin amount. Discuss the
appropriate percentage for each health event with your healthcare professional.