13.
If the correct PIN was entered,
this Pair with Pump screen
appears on the meter, and the
DEVICE PAIRED screen appears
on the pump.
Pump:
Press f to confirm the
j
pairing.
66
Meter:
Turn o the meter. Once the
j
meter has turned o , the
pairing process is complete.
You may turn on the meter and
begin using it.
NOTE
• A blood glucose test cannot
be performed during the
pairing process.
• The meter and pump remain
paired when the batteries
are removed and replaced.
• Only one meter can be
paired with one pump at a
time.
• If the meter is paired with
a pump, and you pair the
meter with another pump,
the meter is unpaired from
the first pump.
• If the meter is paired with a
pump and an unsuccessful
attempt is made to pair the
meter to another pump, the
meter is no longer paired
with the original pump.