System Specifications; Environmental Conditions - ResMed S7 lightweight User Manual

Hide thumbs Also See for S7 lightweight:
Table of Contents

Advertisement

S
S
YSTEM
PECIFICATIONS
Performance
Operating pressure range: 4 to 20 cm H
Dimensions (L x W x H): 10.6" x 9.1" x 5.6"
Weight: 4.6 lb.
Power Supply
Input range 100–240 V, 50–60 Hz, < 120 VA (maximum power consumption)
Actual power consumption will vary, depending on factors such as the use of
accessories, height above sea level, and ambient temperature. Power consumption
values for typical treatment conditions (using a breathing machine set to 0.5 L and 15
breaths/min with 6'6" air tubing and a modular mask) are given in the following table.
Device
ResMed S7 Lightweight
Housing Construction: Flame retardant engineering thermoplastic

Environmental Conditions

Operating Temperature: +41°F (+5°C) to +104°F (+40°C)
Operating Humidity: 10%–95% non-condensing
Storage and Transport Temperature: -4°F (-20°C) to +140°F (+60°C)
Storage and Transport Humidity: 10%–95% non-condensing
Electromagnetic Compatibility: Product complies with all applicable
electromagnetic compatibility requirements (EMC) according to IEC60601-1-2, for
residential, commercial, and light industry environments. For further details, see
"Guidance and Manufacturer's Declaration - Electromagnetic Emissions and Immunity"
on page 25.
Air Filter: Two-layered, powder-bonded, polyester non-woven fiber
Air Tubing: Flexible plastic, 6'6" (2 m) length
IEC 60601-1 Classifications: Class II (double insulation), Type CF
Note: The manufacturer reserves the right to change these specifications without notice.
O
2
Pressure
Power
(cm H
O)
(VA)
2
5
20
10
24
15
29
20
35
S