ResMed S7 lightweight User Manual
Hide thumbs Also See for S7 lightweight:

Advertisement

R
M
ES
ED
U
'
M
SER
S
ANUAL
Reorder number: 308167/1 04 05

Advertisement

Table of Contents
loading

Summary of Contents for ResMed S7 lightweight

  • Page 1 ’ ANUAL Reorder number: 308167/1 04 05...
  • Page 2 Manufactured by: ResMed Ltd 97 Waterloo Road North Ryde NSW 2113 Australia Tel: +61 (2) 9886 5000 or 1 800 658 189 (toll free) Fax: +61 (2) 9878 0120 Email: reception@resmed.com.au Distributed by: ResMed Corp 14040 Danielson Street Poway CA 92064-6857 USA Tel: +1 (858) 746-2400 or 1-800-424-0737 (toll free) Fax: +1 (858) 746-2900 Email: reception@resmed.com...
  • Page 3: Table Of Contents

    ONTENTS ..........1 NTRODUCTION EFINITIONS WNER...
  • Page 5: Ntroduction

    NTRODUCTION EFINITIONS This manual contains special terms that appear in the text to draw your attention to specific and important information. WARNING Alerts you to possible injury. CAUTION Explains special measures for the safe and effective use of the device. Note: Is an informative or helpful note.
  • Page 6: Warnings

    S7 L unit should only be used with masks (and IGHTWEIGHT • connectors) recommended by ResMed, or by your physician or respiratory therapist. A mask should not be used unless the S7 L unit is IGHTWEIGHT turned on and operating properly. The vent hole or holes associated with the mask should never be blocked.
  • Page 7: Cautions

    Explosion hazard—do not use in the vicinity of flammable anesthetics. • Blocking the hose while in operation could lead to overheating of the device. • ARNINGS RELATED TO OXYGEN USE WARNING If oxygen is used with this device, the oxygen flow must be turned off when the device is not operating.
  • Page 8 OSSIBLE FFECTS The R S7 L flow generator helps you have a good night’s sleep. IGHTWEIGHT However, you need to be aware of possible problems that may arise during CPAP treatment. WARNING Consult your clinician immediately if you experience any of the following symptoms during your CPAP treatment: headache •...
  • Page 9 S7 L IGHTWEIGHT OMPONENTS Please identify and familiarize yourself with the following components of the R S7 L unit: IGHTWEIGHT ResMed S7 Lightweight unit Front view Handle screen Control panel Air outlet Removable front cap Air tubing, 6’6" (2 m)
  • Page 10 2i heated humidifier 2iC passover humidifier UMID UMID ResMed P heated humidifier humidifier UMID ASSOVER WARNING Only the HumidAire 2i, HumidAire 2iC, HumidAire heated humidifier, and the ResMed Passover are compatible for use with the ResMed S7 Lightweight. Please...
  • Page 11: The Res Med S7 Lightweight Ystem

    refer to “Warnings” on page 2. S7 L IGHTWEIGHT YSTEM...
  • Page 13: Setting Use Preparing The Res Med S7 Lightweight

    Note: The ResMed S7 Lightweight unit may be placed on the floor beside or underneath your bed. Ensure that the area is free from dust and clear of bedding, clothes, or any other objects that could block the air inlet.
  • Page 14 Notes The ResMed S7 Lightweight has no humidifier setting options. • The ResMed S7 Lightweight does not have the Warm-Up and cooling down features • mentioned in the HumidAire 2i User’s Manual. When the HumidAire 2i is attached to the ResMed S7 Lightweight, you may begin therapy immediately.
  • Page 15 6’6" or 9’10" (2 m to 3 m) to the right connector port on the humidifier. Close the lid. UMID ResMed Passover Users Connect the medium 21in (52 cm) air tubing to the left connector port, and the long air tubing 6’6" or 9’10" (2 m to 3 m) to the right connector port on the humidifier.
  • Page 16 HumidAire Users Connect the free end of the medium air tubing to the air outlet of the R S7 L IGHTWEIGHT ResMed Passover Users Connect the free end of the medium air tubing to the air outlet of the R S7 L IGHTWEIGHT...
  • Page 17 WARNING Make sure that the power cord and plug are in good condition and the equipment is not damaged. The R S7 L is now ready for use with the H or ResMed IGHTWEIGHT UMID ASSOVER LTITUDE OMPENSATION Changes in altitude will affect the pressure delivered by the R...
  • Page 18 IGHTWEIGHT Function Front • Starts or stops treatment Up/Down • Allows you to scroll through the ResMed S7 L menus IGHTWEIGHT and setting options. • Performs the function indicated by the guiding text displayed Left above it on the LCD screen.
  • Page 19 ALT: 0-2000ft change exit USED HRS: 8888 exit Figure 1: ResMed S7 Lightweight Menu Series UNCTIONS The R S7 L menu functions are summarized in Table 1 with a brief IGHTWEIGHT description of what each function does and the available setting options. To access these functions see “Using the Menus”...
  • Page 20 Table 1: Settings Menu Functions Function Function Description Setting Options Altitude Sets the altitude compensation 0–2000 ft (0–610 m), (Compensation) range. 2001–4000 ft (611–1219 m), 4001–7000 ft (1220–2134 m) Used hours Displays the total number of (View only) treatment hours.
  • Page 21: Operating Instructions

    IGHTWEIGHT tubing and mask system connected. See “Setting Up The ResMed S7 Lightweight” on page 9. Turn the main power switch at the back of the unit to on (I).
  • Page 22: Helpful Hints

    ELPFUL INTS TARTING OUT IRST When you first use your R S7 L unit, you may feel uncomfortable IGHTWEIGHT breathing against the air flow. This is normal. You will continue to breathe normally while you sleep. Take time to adjust to this new sensation. Taking a few deep breaths may help.
  • Page 23: Cleaning And Maintenance

    • scented oils), moisturizing, or antibacterial soaps to clean the cushion, mask, air tubing, or the ResMed S7 Lightweight. These solutions may cause hardening and reduce the life of the product. Do not wash or dry the mask frame at a temperature above 176 F (80 •...
  • Page 24: Replacing The Air Filter

    CAUTION Do not attempt to open the ResMed S7 Lightweight. There are no user serviceable parts inside. Repairs and internal servicing should only be performed by an authorized service agent. EPLACING THE ILTER Inspect the air filter every month to check if it is blocked by dirt or contains holes. With...
  • Page 25: Troubleshooting

    ROUBLESHOOTING If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the unit. Problem Possible Cause Solution No display. Power not connected or Ensure the power cable is switch at back is not on.
  • Page 27: System Specifications

    6’6" air tubing and a modular mask) are given in the following table. Pressure Power Device (cm H (VA) ResMed S7 Lightweight Housing Construction: Flame retardant engineering thermoplastic Environmental Conditions Operating Temperature: +41°F (+5°C) to +104°F (+40°C) Operating Humidity: 10%–95% non-condensing Storage and Transport Temperature: -4°F (-20°C) to +140°F (+60°C) Storage and Transport Humidity: 10%–95% non-condensing...
  • Page 28 LOSSARY OF YMBOLS Attention, consult accompanying documents Class II equipment Type CF equipment Drip Proof Start/Stop RESSURE HARACTERISTICS Pressure (cm H Vista/Mirage/Protégé Modular Mirage Full Face Mask/Mirage Full Face Mask Series 2 Ultra Mirage...
  • Page 29 Guidance and manufacturer’s declaration – electromagnetic emissions The ResMed S7 Lightweight is intended for use in the electromagnetic environment specified below. The customer or the user of the ResMed S7 Lightweight should assure that it is used in such an environment. Electromagnetic environment -...
  • Page 30 Guidance and manufacturer’s declaration – electromagnetic immunity The ResMed S7 Lightweight is intended for use in the electromagnetic environment specified below. The customer or the user of the ResMed S7 Lightweight should assure that it is used in such an environment. IEC60601-1-2...
  • Page 31 Guidance and manufacturer’s declaration – electromagnetic immunity The ResMed S7 Lightweight is intended for use in the electromagnetic environment specified below. The customer or the user of the ResMed S7 Lightweight should assure that it is used in such an environment. IEC60601-1-2...
  • Page 32 Recommended separation distances between portable and mobile RF communications equipment and the ResMed S7 Lightweight The ResMed S7 Lightweight is intended for use in an environment in which radiated RF disturbances are controlled. The customer or the user of the ResMed S7 Lightweight can help prevent electromagnetic interference...
  • Page 33: Limited Warranty

    Note: Some models are not available in all regions. If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components. This Limited Warranty does not cover: a) any damage caused as a result of improper use, abuse, modification or alteration of the product;...
  • Page 35: Index

    Operating Instructions 17 Electromagnetic Compatibility 23 Environmental Conditions 23 Error Messages 21 Performance 23 Power Supply 23 Preparing for Use 9 Features of the ResMed S7 Lightweight 14 First Use 18 Flow/Pressure Characteristics 24 Ramp Front Key 14 screen 14 time 14...
  • Page 36 Travelling 18 Troubleshooting 21 Up/Down Key 14 User/Owner Responsibility 1 Using the Menus 14 Warnings Related to Oxygen Use 3 Warnings Related to Treatment 3 Weekly Cleaning 19 Weight 23...

Table of Contents