Introduction; Required Equipment - Fluke 8845A Calibration Manual

Digital multimeter
Hide thumbs Also See for 8845A:
Table of Contents

Advertisement

Introduction

This chapter of the Calibration Manual provides performance tests to verify the Meter is
operating within published specifications as well as a complete calibration procedure.
The performance test and, if necessary, the calibration procedure can be performed both
periodically and after service or repair.
The performance tests can be used as an acceptance test upon receipt of the Meter. Use
the 90-day specifications when performing an acceptance test after performing a
calibration.

Required Equipment

Table 3-1 lists the equipment required for performance testing and calibration of the
Meter.
Function
Instrument Type
Volts dc
Standard
8½ digit meter
4-wire short
Alternate standard
Volts ac
Standard
8½ digit meter
4-wire short
Alternate standard
Frequency
Standard
Alternate standard
Alternate standard
Table 3-1. Required Test Equipment
Model
Fluke 5520A
Fluke 8508A
Fluke low thermal 4-wire
short or equivalent
Fluke 5720A
Fluke 5520A
Fluke 8508A
Fluke low thermal 4-wire
short or equivalent
Fluke 5720A
Fluke 5520A
Fluke 5520A with any
scope option
Function generator
Performance Test and Calibration
Comments
Must be characterized with 8508A
Used to characterize the 5520A
Fluke PN 2653346
Must be characterized with 8508A
Used to characterize the 5520A.
Note: TURs <4:1 at 1 V, 10 V, and
100 V at 20 kHz
Fluke PN 2653346
Specifications include0.075 %
frequency accuracy from 3 – 40 Hz
and 0.0025 % accuracy for
frequencies up to 1 MHz
3

Introduction

3-3