Page 2
8808A Calibration Manual Supplement Change #1 On page 3-11, replace Table 3-8 with the following: Table 3-8. Frequency Verification Steps Nominal Input 90-Day Test Limit 1-Year Test Limit Range Ampl. Freq. High High 0.1 V 20 Hz 2000 Hz 20.04 Hz 19.96 Hz...
Page 3
Manual Supplement 8808A Calibration Change #3, 64137, 64189 On page 1-4, add the following to the Symbols, table: Conforms to relevant South Korean EMC Standards. Conforms to CAN/CSA-C22.2 No. 61010-1, second edition, including Amendment 1. Electromagnetic On page 1-9, add...