Download Print this page
Fluke 8808A Manual Supplement
Hide thumbs Also See for 8808A:

Advertisement

Quick Links

Manual Title:
Print Date:
Revision/Date:
This supplement contains information necessary to ensure the accuracy of the
above manual. This manual is distributed as an electronic manual on the
following CD-ROM:
.
© 2010-2012 Fluke Corporation. All rights reserved.
Manual Supplement
8808A Calibration
October 2008
CD Title:
CD Rev. & Date:
CD PN:
Supplement Issue:
Issue Date:
Page Count:
8808A
3, 12/2009
2713002
4
12/12
2

Advertisement

loading

Summary of Contents for Fluke 8808A

  • Page 1 This supplement contains information necessary to ensure the accuracy of the above manual. This manual is distributed as an electronic manual on the following CD-ROM: CD Title: 8808A CD Rev. & Date: 3, 12/2009 CD PN: 2713002  © 2010-2012 Fluke Corporation. All rights reserved. ...
  • Page 2 8808A Calibration Manual Supplement Change #1 On page 3-11, replace Table 3-8 with the following: Table 3-8. Frequency Verification Steps Nominal Input 90-Day Test Limit 1-Year Test Limit Range Ampl. Freq. High High 0.1 V 20 Hz 2000 Hz 20.04 Hz 19.96 Hz...
  • Page 3 Manual Supplement 8808A Calibration Change #3, 64137, 64189 On page 1-4, add the following to the Symbols, table:  Conforms to relevant South Korean EMC Standards. Conforms to CAN/CSA-C22.2 No. 61010-1, second edition,  including Amendment 1. Electromagnetic On page 1-9, add...