Fluke 233 - True-rms Remote Display Digital Multimeter

Fluke 233 - True-rms Remote Display Digital Multimeter

Also See for 233:

Advertisement

Fluke 233 - True-rms Remote Display Digital Multimeter

Introduction

The Fluke 233 (hereafter the Meter) is a compact and easy to operate tool for electrical and electronic circuit measurements.


Read "Safety Information" before you use the Meter.

Safety Information

The Meter complies with:

  • ISA-82.02.01
  • CAN/CSA C22.2 No. 61010-1-04
  • ANSI/UL 61010-1:2004
  • EN 61010-1:2001
  • EN 61326-1:2006
  • EN 61326-2-2:2006
  • ETSI EN 300 328 V1.7.1:2006
  • ETSI EN 300 489 V1.8.1:2008
  • FCC Part 15 Subpart C Sections 15.207, 15.209, 15.249 FCCID: T68-F233
  • RSS-210 IC: 6627A-F233
  • Measurement Category III, 1000V, Pollution Degree 2
  • Measurement Category IV, 600V, Pollution Degree 2

In this manual, a Warning identifies the conditions and procedures that cause a dangerous situation to the user. A Caution identifies the conditions and procedures that could cause damage to the Meter, equipment under test damage, or permanent data loss.

The symbols used on the Meter and in this manual are shown in Table 1.


To prevent possible electrical shock or personal injury, follow these guidelines:

  • Use this Meter only as specified in this manual or the protection can be compromised.
  • Do not use the Meter if it is damaged. Before you use the Meter, examine the case. Look for cracks or missing plastic. Carefully look at the insulation around the terminals.
  • Make sure the battery door is closed and locked before you operate the Meter.
  • Replace the batteries when the battery indicator () appears.
  • Remove the test leads from the Meter before the battery door on the Meter base is opened.
  • Examine the test leads for damaged insulation or exposed metal. Measure the test leads for continuity. Replace damaged test leads before you use the Meter.
  • Do not apply more than the rated voltage, shown on the Meter, between the terminals or between a terminal and earth ground.
  • Do not operate the Meter with the battery door removed or the case open.
  • Be careful around voltages >30 V ac rms, 42 V ac peak, or 60 V dc. These voltages pose a shock hazard.
  • Use only the replacement fuse specified by the manual.
  • Use the correct terminals, function, and range for measurements.
  • Do not work alone.
  • For current measurements, connect the Meter to the circuit after you remove circuit power. Always put the Meter in series with the circuit.
  • Connect the common test lead before the live test lead and remove the live test lead before the common test lead.
  • Do not use the Meter if it operates incorrectly. Protection can be compromised. If you are unsure, have the Meter examined.
  • Do not use the Meter around explosive gas, vapor or in damp or wet environments.
  • Use only specified 1.5-V AA batteries (three in the Meter base and two in the display), correctly installed, for Meter power.
  • Comply with local and national safety requirements when in hazardous locations.
  • Only use test leads that have the same voltage, category, and amperage ratings as the Meter and that are approved by a safety Agency.
  • Measure a known voltage first to make sure that the Meter operates correctly. If you are unsure, have the Meter examined.
  • Use protective equipment, as directed by local or national authorities when in hazardous work areas.
  • Measure the test leads for continuity before use. Do not use if the resistance is high or noisy.
  • Use only specified replacement parts in the Meter.
  • Keep fingers behind the finger guards on the probes.


To prevent damage to the Meter or to the equipment under test, follow these guidelines:

  • Disconnect circuit power and discharge all high-voltage capacitors before you do diode tests or measure resistance, continuity, or capacitance.
  • Use the correct terminals, function, and range for all measurements.
  • Before a current measurement, do the fuse test.

Table 1. Electrical Symbols

AC (Alternating Current) Earth ground
DC (Direct Current) Fuse
shock hazard Hazardous voltage Conforms to European Union directives.
warning Risk of Danger. Important information. See Manual. Conforms to relevant Canadian Standards Association directives.
Battery. Low battery when shown. Double insulated
Continuity test or continuity beeper tone. Capacitance
CAT III IEC Measurement Category III
CAT III equipment has protection against transients in equipment in fixedequipment installations, such as distribution panels, feeders and short branch circuits, and lighting systems in large buildings.
CAT IV

IEC Measurement Category IV

CAT IV equipment has protection against transients from the primary supply level, such as an electricity meter or an overhead or underground utility service.

Do not discard this product as unsorted municipal waste. Go to the Fluke website for recycling data. Diode
Examined and licensed by TÜV Product Services. Conforms to relevant Australian standards.

Features

See Tables 3 through 4 for a list of Meter features with a short feature description.

Table 2. Display

Display
No. Symbol Indication
1 MIN MAX AVG mode on.
2 MAX MIN AVG Maximum, minimum, or average measurement shown.
3 Display hold on. Display freezes the measurement.
4 Radio connection indicator.
5 °C, °F degrees Celsius, degrees Fahrenheit
6 A amperes (amps)
V, mV volts, millivolts
μF, nF microfarad, nanofarad
DC
AC
Direct current or alternating current.
Ω, MΩ, kΩ ohm, megohm, kilohm
Hz, kHz hertz, kilohertz
7 remote Battery low warning for the display module.
8 Manual Range Manual range set.
9 Auto Range Auto range set.
10 meter Battery low warning for Meter base.
11 Diode test mode.
12 Continuity test.
13 Input is a negative value.
14 shock hazard Hazardous voltage. Measured input voltage ≥30 V, or voltage overload condition (OL). For frequency measurements >1 kHz, the  symbol and high-voltage LED is unspecified.

Table 3. Inputs

 Inputs
No. Terminal Description
1 A Input for 0 A to 10.00 A current measurements.
2 COM Common terminal for all measurements.
3 Input for voltage, continuity, resistance, diode, capacitance, temperature, and frequency measurements.

Table 4. Function Switch Positions

Switch Position Description

Hz (button)
AC voltage from 0.06 to 1000 V.
Frequency from 5 Hz to 50 kHz.
DC voltage from 0.001 V to 1000 V.
AC voltage from 6.0 to 600.0 mV, dc-coupled. DC voltage from 0.1 to 600.0 mV.
Ohms from 0.1 Ω to 40 MΩ.
Continuity beeper turns on at <20 Ω and turns off at >250 Ω.
Farads from 1 nF to 9999 μF.
Diode Test. OL shows in the display when the input voltage is >2.0 V.
Temperature.

Hz (button)

AC current from 0.1 A to 10 A (>10 to 20 A, 30 seconds on, 10 minutes off). >10.00 A display flashes. >20 A, OL is shown. DC-coupled.

Frequency from 45 Hz to 5 kHz.

DC current from 0.001 A to 10 A (>10 to 20 A, 30 seconds on, 10 minutes off). >10.00 A display flashes. >20 A, OL is shown.
warning Note: All ac functions are true-rms. AC voltage is ac-coupled. AC mV and ac amps are dc-coupled.

Error Messages

Table 5 contains possible error messages and the steps to clear the error.

Table 5. Error Messages

Error Messages
Display-module batteries must be replaced before the Meter will operate.
Meter-base batteries must be replaced before the Meter will operate.
Calibration necessary. Meter calibration is necessary before the Meter will operate.
Internal error. The Meter must be repaired before it will operate.
Loss of radio connection with the Meter base.

Remote Operation

Figure 1. Display Module Separation
Display Module Separation

Figure 2. Dock Display Module with Meter Base
Dock Display Module with Meter Base

Fuse Test

Figure 3. Fuse Test
Fuse Test

Battery Replacement

Figure 4. Display-Module Battery Removal
Display-Module Battery Removal

Figure 5. Meter Base Battery Replacement
Meter Base Battery Replacement

General Specifications

Maximum voltage between any terminal and earth ground: 1000 V rms

warning Fuse for A inputs: 11 A, 1000 V 17000A interrupt rating Fuse

Altitude

Operating: 2,000 meters
Storage: 12,000 meters

Temperature

Operating: -10°C to +50°C
Storage: -40°C to +60°C

Temperature coefficient: 0.1 X (specified accuracy) / °C (< 18°C or > 28°C)

Electromagnetic Compatibility (EN 61326-1:2006):
In an RF field of 3 V/m, accuracy = specified accuracy except in temperature: specified accuracy ±5°C (9°F)

Wireless Frequency: 2.4 GHz ISM Band 10 meter range

Relative Humidity: Maximum noncondensing, 90% at 35°C, 75% at 40°C, 45% at 50°C, 0% to 70% for 40 MΩ range

Battery Type

Meter base: Three AA Alkaline batteries, NEDA 15A IEC LR6
Display module: Two AA Alkaline batteries, NEDA 15A IEC LR6

Battery Life: 400 hrs typical (Alkaline)

Shock: 1 Meter drop 6 sides per IEC 61010

Safety Compliance:

Complies with ANSI/ISA S82.01-2004, CSA 22.2 No. 61010-1-04 to 1000 V
Measurement Category III and 600 V Measurement Category IV.

Certifications: CSA, TÜV (EN61010), ; (N10140),VDE, GOST

LIMITED WARRANTY AND LIMITATION OF LIABILITY

This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke's behalf. To obtain service during the warranty period, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that Service Center with a description of the problem.

THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.

Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.

Fluke Europe B.V.
P.O.Box 1186
5602 BD Eindhoven
The Netherlands

How to Contact Fluke

Use one of the telephone numbers below to speak with a Fluke representative:

Technical Support USA: 1-800-44-FLUKE (1-800443-5853)
Calibration/Repair USA: 1-888-99-FLUKE (1-888- 993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500 Or, visit Fluke's website at www.fluke.com.
To register your product, visit http://register.fluke.com.

Documents / Resources

References

Download manual

Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.

Download Fluke 233 - True-rms Remote Display Digital Multimeter

Advertisement

Need Assistance?

Do you have a question about the 233 that isn't answered in the manual? Leave your question here.

Error while submitting your question. Please, reload page and try again.

Thank you! Your question has been received!

Table of Contents