Dc Current Verification Steps - Fluke 8845A Calibration Manual

Digital multimeter
Hide thumbs Also See for 8845A:
Table of Contents

Advertisement

8845A/8846A
Calibration Manual

DC Current Verification Steps

Connect the Meter to the test equipment as shown in Figure 3-7 and, depending on which
meter you are calibrating, apply the nominal values listed in Table 3-14 or Table 3-15.
Verification forms can be found in Appendix A which can be copied and used to record
each meter reading.
3-20
INPUT
SENSE
V
4 W
2W/4W
HI
HI
1000 V CAT I
300V
600V CAT II
LO
LO
1V
FUSED
100
10A
mA
REAR
FRONT
5520A
CALIBRATOR
NORMAL
AUX
V, , ,RTD
A, -SENSE, AUX V
HI
LO
GUARD
20A
TC
20V PK MAX
20V PK MAX
UUT
A.
EX SNS
INPUT
SENSE
V
4 W
2W/4W
HI
HI
1000 V CAT I
300V
600V CAT II
LO
LO
1V
FUSED
100
10A
mA
REAR
FRONT
NC
Figure 3-7. 100 mA DC Current Equipment Setup
8846A
6-1/2 DIGIT PRECISION MULTIMETER
+100.0031
mA
BACK
F1
F2
F3
F4
DC V
DC I
AC I
TRIG
ACV
RANGE
FREQ
TEMP
ZERO
ANALYZE
PERIOD
SCOPE
OUT
STBY
OPR
EARTH
SCOPE
μ
7
8
9
TRIG
n
4
5
6
p
1
2
3
/ +
0
SHIFT
: OFF
: OFF
EX GRD
F5
MEMORY
INSTR
SETUP
MEAS
SETUP
PREV
EXGRD
MENU
dBm
sec
m
V
Hz
W
F
k
A
C
M
F
ENTER
5720A Calibrator
OUTPUT
SENSE
V A
V
WIDEBAND
HI
HI
LO
LO
HI
AUX
GUARD
GROUND
CURRENT
caw022.eps

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8846a

Table of Contents