Eizo SCD 19102 Instruction Manual page 67

High end 19” color and gray scale flat panel display
Hide thumbs Also See for SCD 19102:
Table of Contents

Advertisement

4. Execute "Force"
Optimize scanning frequency
The correct signal frequency must now be set.
1. Select Mode 4 from line 6 in the "Live mode" menu using button 2.
2. Connect complete SMPTE test pattern: Optimize the signal frequency in the "Total
horizontal lines" input field so that the squares of the test pattern are approximately
square.
If the squares are stretched horizontally (top and bottom edges are longer than the
sides), the value "Total horizontal lines" should be reduced.
If the squares are compressed (top and bottom edges are smaller than the sides), the
value "Total horizontal lines" should be increased.
Note
Setting limit reached
• Since the resolution and scanning frequency can only be set within limits, the "Force"
It may be necessary to repeat this process several times!
• The value "Total horizontal lines" must always be greater than the value "Horizontal
SCD 19102 / SMD 19102
Instruction Manual, 05/2008, A5E01588521A-002
function must be carried out when the setting limit is reached (selected using Mode 4
from line 6 in the "Live mode" menu) in order to continue with the setting process. In
this manner, the slider for the setting range is again set in the center.
resolution". If, when setting the "Total horizontal lines", the "Horizontal resolution" is
greater, it should be reduced. A larger "Horizontal resolution" can result in the image
being split vertically.
command using key 2.
Commissioning
7.4 Make settings
67