Application planning

4.1

Installation site

Provide adequate ventilation
Ventilation slots are located on the rear of the housing.
The distance at the rear must be at least 10 cm from a wall or 15 cm from other units.
Observe the permissible ambient temperature range
The unit must not be operated outside the permissible ambient temperature range.
Desktop installation
If the SCD / SMD 19102 is mounted on a stand, it must be positioned on a hard, horizontal
surface.
Avoid reflections on the screen
The display has an anti-glare surface which is only effective with a clean, grease-free
screen.
The SCD / SMD 19102 CP display includes an optically coated protective glass pane, which
is well suited for use in a germ-free environment when cleaning agents are used.
If the screen surface is dirty, clean it using a suitable microfiber cloth.
Please note the additional cleaning instructions, see the section "Service and
Maintenance" (Page 73).
The display should be positioned so that reflections of lights, windows, furniture with shiny
surfaces or light-colored walls do not appear on the screen.
In order to reduce mirroring on the unit, only use non-dazzling reflector lights for the ceiling
lighting.
Change of environment
If the unit is brought into a warm environment from a cold one, condensation may form in the
unit. The unit should not be switched on until all the condensed water has evaporated,
including that inside the unit. This may take several hours, depending on the conditions.
See also
Cleaning (Page 73)
SCD 19102 / SMD 19102
Instruction Manual, 05/2008, A5E01588521A-002