ResMed Defy Mini User Manual page 36

Clinical guide
Table of Contents

Advertisement

 
Physical - device and humidifier
Dimensions (H x W x D):
Air outlet (complies with ISO 5356-1:2004):
Weight (device and cleanable humidifier):
Housing construction:
Water capacity:
Cleanable humidifier - material:
Temperature
Maximum heater plate:
Cut-out:
Maximum gas temperature:
Air filter
Standard:
Hypoallergenic:
Aircraft use
ResMed confirms that device meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21,
category M) for all phases of air travel.
Wireless module
Technology used:
It is recommended that the device is a minimum distance of 2 cm from the body during operation. Not applicable to masks,
tubes or accessories
Declaration of Conformity (DoC to the R&TTE Directive)
ResMed declares that the AirCurve 10 device is in compliance with the essential requirements and other relevant
provisions of Directive 1999/5/EC. A copy of the declaration of conformity (DoC) can be found on
www.ResMed.com/ProductSupport.
Operating pressure range
ASV, ASVAuto:
CPAP
Supplemental Oxygen
Maximum flow:
Pneumatic flow path
Design life
Device, power supply unit:
Cleanable humidifier:
Air tubing:
116 mm x 255 mm x 150 mm
22 mm
1248 g
Flame retardant engineering thermoplastic
To maximum fill line 380 mL
Injection moulded plastic, stainless steel and silicone seal
68°C
74°C
 41°C
Material: Polyester non woven fibre
Average arrestance: >75% for ~7 micron dust
Material: Acrylic and polypropylene fibres in a polypropylene
carrier
Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron
dust
2G GSM
4 to 25 cm H
O (4 to 25 hPa)
2
4 to 20 cm H
O (4 to 20 hPa)
2
15 L/min (CPAP, ASV); 4 L/min (ASVAuto)
1. Flow sensor
2. Blower
3. Pressure sensor
4. Mask
5. Air tubing
6. Humidifier
7. Device
8. Inlet filter
5 years
2.5 years
6 months
English
33

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Aircurve 10 pacewave

Table of Contents