Humidifier performance
The following settings have been tested at 22°C ambient temperature:
Mask Pressure
cm H
O (hPa)
2
3
4
10
20
25
1
AH - Absolute Humidity in mg/L
2
BTPS - Body Temperature Pressure Saturated
Air tubing
Air tubing
Material
ClimateLineAir
Flexible plastic and electrical components
SlimLine
Flexible plastic
Standard
Flexible plastic
Heated air tubing temperature cut-out: 41°C
Notes:
The manufacturer reserves the right to change these specifications without notice.
The electrical connector end of the heated air tubing is only compatible with the air outlet at the device end and should
not be fitted to the mask.
Do not use electrically conductive or antistatic air tubing.
The temperature and relative humidity settings displayed are not measured values.
Displayed values
Value
Pressure sensor at air outlet:
Mask pressure
Flow derived values:
Leak
Tidal volume
Respiratory rate
Minute ventilation
Value
Pressure measurement:
2
Mask pressure
Flow and flow derived values
Flow
2
Leak
2,3
Tidal volume
2,3
Respiratory rate
2,3
Minute ventilation
1
Results are expressed at ATPD (Ambient Temperature and Pressure, Dry).
2
Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal volumes <100 mL or minute ventilation <3 L/min.
3
Measurement accuracy verified as per EN ISO 10651-6:2009 for Home Care Ventilatory Support Devices (Figure 101 and Table 101) using
nominal ResMed mask vent flows.
34
RH output %
Setting 4
85
85
85
85
85
Range
4–25 cm H
0–120 L/min
0–4000 mL
0–50 BPM
0–30 L/min
Accuracy
±[0.5 cm H
1
:
±6 L/min or 10% of reading, whichever is greater, at 0 to 150 L/min positive flow
±12 L/min or 20% of reading, whichever is greater, 0 to 60 L/min
±20%
±1.0 BPM
±20%
Setting 8
100
100
100
90
90
O (4–25 hPa)
2
1
O (0.5 hPa) + 4% of measured value]
2
Nominal system output AH
Setting 4
6
6
6
6
6
Length
2 m
1.8 m
2 m
Display resolution
0.1 cm H
O (0.1 hPa)
2
1 L/min
1 mL
1 BPM
0.1 L/min
1
2
, BTPS
Setting 8
>10
>10
>10
>10
>10
Inner diameter
15 mm
15 mm
19 mm